Gather the following items:
Make sure you have the code chip
ready to insert if you are opening
• Coaguchek XS Pro Meter
a new box of strips. The code key
• Container of Test Strips
needs to be inserted into the meter
• Test Strip Code Chip
everytime you use a new lot of strips.
• Safe-T-Pro Plus Lancet
• Alcohol Wipe
• Cotton Ball or Tissues
7.
8.
Touch the Scan button and hold the
Press Patient Test on the main menu
operator ID barcode approximately
screen.
10cm from the built-in scanner. The
meter beeps once the barcode has
been successfully read.
0508 MY LIFE | 0508 69 54 33
For Assistance Phone
Quick Reference Guide
3.
4.
Clean the fingertip with an alcohol
wipe. Or have the patient wash his or
her hands in warm soapy water.
Dry thoroughly.
9.
10.
Touch the Scan button and hold the
patient ID barcode approximately
10cm from the built-in barcode
scanner. The meter beeps once the
barcode has read successfully.
An hour glass symbol appears.
The test strip symbol appears after a
few seconds with a flashing drop of
blood. You have 180 seconds to apply
the sample.
5.
Before continuing, review these
Take a test strip out of the container.
tips for getting a good drop of
Close the container tightly.
blood.
Increasing the blood flow in the
You have 10 minutes to use the test
finger will help you to get a good
strip once you remove it from the
drop of blood so keep in mind these
container.
tips:
•
Warm the hand. Have the patient
hold it under his or her arm, and/
or wash with warm water.
•
Have the patient let that arm
hang by his or her side.
•
Massage the finger from its base.
Use these techniques until the
fingertip has increased in colour.
11.
Twist the protective cap off the
Within 15 seconds of lancing the fingertip,
lancet.
apply the blood to the target area on the strip.
Find the clear target area on the test strip. You
can dose from the side or the top.
Massage the finger until you see
increased colour in the fingertip.
Hold the blood drop to the test strip until you
hear a beep. The flashing blood drop symbol
will disappear and an hourglass symbol appears
Keeping the hand down, press the tip
and the test begins.
of the lancet firmly against the side of
Do not add more blood to the test strip.
the fingertip. Press the purple trigger
Do not touch the test strip while the test
button.
is in progress.
The meter performs an automatic quality
Gently squeeze from the base of the
control check on the strip before it displays
finger to develop a drop of blood.
the result.
QC √ appears in the display briefly following a
successful QC check. The result appears in less
than a minute.
6.
Slide the test strip into the test strip
guide in the direction of the arrows
until it stops.
The meter turns on and the main
menu appears.
If you insert a strip from a new lot
number for the first time, the meter
will display the number of the test
strip code chip required. Remove
the old code chip and insert the new
one.
12.
Record your result as instructed by
your facility.
Note: If during testing the meter
displays an error message, refer to
the Error Messages section of the
user manual for an explanation and
steps on how to proceed.