4. Remove the cap from the one-time-use 21g lancet,
place it against the skin, and push down firmly against
the finger to lance the surface of skin. Note: smaller
gauge lancets should not be used.
The blood should bleed freely. If it doesn't, gently
5.
squeeze the finger on each side of the wound to
get it started. Lower your hand below your heart if
necessary to produce a "bead" sized drop of blood.
Hold sample transfer tube below bulb. DO NOT
6.
TOUCH OR SQUEEZE BULB. With tube horizontal
touch tip to bead of blood. Let capillary action fill cap-
illary portion of tube until blood flow stops. Squeeze
finger to generate additional blood if required to com-
pletely fill capillary portion of sample transfer tube.
7. Move fingers to flat sides of bulb being sure to cover
air hole. Rest hand on meter. Insert tip into sample
application well of test strip, touching tip down on
strip at flashing green light. Gently squeeze bulb until
blood leaves tube (do not produce air bubbles). Keep
pressure on bulb and pull away from test strip (avoids
back suction). You must apply blood to the test
strip within 15 seconds of lancing the finger.
The meter will know that you have added the sample
and the display will then read: "TESTING, PLEASE
WAIT". Testing time is patient's actual clotting time.
When testing is complete, the meter beeps once
8.
and the results (INR and Prothrombin Time (PT) in
seconds) appear on the screen.
Record the results. Remove the test strip and properly
dispose along with the transfer tube and lancet. Turn
off the meter.
NOTE: Read Self-Test User's Manual for complete instructions for use, indications, warnings, precautions and
troubleshooting tips. If any error messages are displayed, refer to the "Troubleshooting" section of the User
Manual or call CoaguSense Technical Service at 1-866-903-0890.
For In Vitro Diagnostic Use
CSI P/N 100287 Rev. AE