6. Trouble-shooting

Trouble Phenomenon
The instrument does
not show, and does not
work.
The instrument doesn't
work but with show.
The instrument shows
unstable or the
measured value has a
great deviation.
Trouble Reason
Power supply error.
Wiring error.
The sensor doesn't aim at the liquid or the
material.
The surface has great fluctuations.
Liquid surface with lots of foam.
The container is empty and the bottom is not
flat.
Over the range.
The level enters the blanking distance.
There is strong electromagnetic interference
There is Obstruction of the ultrasonic wave.
İMES Sanayi Sitesi B Blok 204. Sk. No:40
Y. Dudullu Ümraniye - İstanbul
Tel:0216 540 7300 Faks: 0216 540 7303
www.aktek.com.tr [email protected]
www..aktek.com.tr
Adjust the sensor and aim at the material.
instrument. or prevent the level too high.
15
Solution
Check the power supply.
Check the wiring.
Add a tube to the container.
Add a tube to the container
Use instrument with a lager range.
Use instrument with a lager range.
Increase the installation height of the
Increase shielding to the instrument.
Change the installation site or using a
plastic tube.