Aktek AT300 Manual do utilizador - Página 13

Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição Aktek AT300. Aktek AT300 16 páginas. Ultrasonic thickness gauge

4.9 Data Printing

At the end of the inspection process, or end of the day, the user may require the
readings be transferred to a computer. The following steps outline this procedure. This
function is only available for AT300, and not for other models.
1.
Before printing, please insert one connection plug of the print cable (Optional parts) into
the socket on the up-left of the main body, and insert the other plug into the
communication socket of the mini-printer.
2.
Press the
key to activate the data logging functions.
3.
Use the
key and the
4.
Press the
key to print the selected file. This operation will send all the data in
current file to the mini printer via RS232 port and print them out.
5.
Press the
key to exit the data logging functions and return to measurement mode.

4.10 Beep Mode

When the beep is set to 【On】 ,it would make a short hoot while press the key each
time, on each measurement, or the measured value exceeds the tolerance limit.
Press the
key to switch the beep mode on and off. It will display the current beep
mode on the main screen.

4.11 EL Backlight

With the background light, it is convenient to work in the dark condition. Press key
to switch on or switch off the background light at any moment as you need after power on.
Since the EL light will consume much power, turn on it only when necessary.

4.12 Battery Information

Two AA size alkaline batteries are needed as the power source. After several hours'
usage of the preset batteries, the battery symbol on the screen will be shown as
The more of dark part indicates the more close to fill. When the battery capacity runs out,
the battery symbol will be shown as
batteries should be replaced.
Please take out the batteries when not working during a long period of time.

4.13 Auto Power Off

The instrument features an auto power off function designed to conserve battery life. If
the tool is idle for 5 minutes, it will turn itself off. While the voltage of the battery is too low
this function will also work.

4.14 System Reset

Press down the
All the memory data will be cleared during system reset. The only time this might possibly
helpful is if the parameter in the gauge was somehow corrupted.

4.15 Connecting to a Computer

AT300 is equipped with a RS232 serial port. Using the accessory cable, the gauge has
the ability to connect to a computer, or external storage device. Measurement data stored in
the memory of the gauge can be transferred to the computer through the RS232 port.
Detailed information of the communication software and its usage refer to the software
key to select the desired file.
and will begin to flash. When this occurs, the
key while powering on the instrument will restore factory defaults.
www.aktek.com.tr
.
13