4.3 Sound Velocity Calibration

In order for the gauge to make accurate measurements, it must be set to the correct
sound velocity for the material being measured. Different types of material have different
inherent sound velocities. If the gauge is not set to the correct sound velocity, all of the
measurements the gauge makes will be erroneous by some fixed percentage. The
One-Point calibration is the simplest and most commonly used calibration procedure
optimizing linearity over large ranges. The Two-point calibration allows for greater accuracy
over small ranges by calculating the probe zero and velocity.
Note: One and Two point calibrations must be performed on material with the paint or
coating removed. Failure to remove the paint or coating prior to calibration will result in a
multi material velocity calculation that may be different from the actual material velocity
intended to be measured.
4.3.1 Calibration to a known thickness
Note: This procedure requires a sample piece of the specific material to be measured, the
exact thickness of which is known, e.g. from having been measured by some other means.
1) Perform a Probe-Zero.
2) Apply couplant to the sample piece.
3) Press the transducer against the sample piece, making sure that the transducer sits flat
against the surface of the sample. The display should show some thickness value, and
the coupling status indicator should appear steadily.
4) Having achieved a stable reading, remove the transducer. If the displayed thickness
changes from the value shown while the transducer was coupled, repeat step 3.
5) Press the
key to activate the calibration mode. The MM (or IN) symbol should begin
flashing.
6) Use the
key and the
matches the thickness of the sample piece.
7) Press the
key again. The M/S (or IN/μS) symbols should begin flashing. The gauge
is now displaying the sound velocity value it has calculated based on the thickness
value that was entered.
8) Press the
key once again to exit the calibration mode and return to the
measurement mode. The gauge is now ready to perform measurements.
4.3.2 Calibration to a known velocity
Note: This procedure requires that the operator knows the sound velocity of the material to
be measured. A table of common materials and their sound velocities can be found in
Appendix A of this manual.
1) Press the
key to activate the calibration mode. The MM (or IN) symbol should begin
flashing.
2) Press the
key again, so that The M/S (or IN/μS) symbols are flashing.
3) Use the
key and the
matches the sound velocity of the material to be measured. You can also press the
key to switch among the preset commonly using velocities.
4) Press the
key to exit from the calibration mode. The gauge is now ready to perform
key to adjust the displayed thickness up or down, until it
key to adjust the sound velocity value up or down, until it