15. EMC and RF statements
AKTIIA bracelet needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided in the following section.
AKTIIA bracelet is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
AKTIIA bracelet uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
AKTIIA bracelet is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. Interference may occur in the vicinity of equipment marked with the following symbol
Portable and mobile RF communication equipment (e.g. cell phones) can affect AKTIIA bracelet.
WARNING: AKTIIA bracelet should not be used adjacent to or stacked with other equipment and if adjacent or
stacked use is necessary, it should be observed in order to verify normal operation in the configuration in which
it will be used.
WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity
of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of AKTIIA bracelet, including cables speci-
fied by the manufacturer. Otherwise, degradation of the performance of this equipment could result."
16. Electromagnetic Compatibility Information
Guidance and manufacture's declaration - electromagnetic emissions
AKTIIA bracelet is intended for use in the electromagnetic environment specified below.
The user of AKTIIA bracelet should ensure that it is used in such an environment.
AKTIIA bracelet is suitable for use in "Home Healthcare
tic establishments and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Emission test
Conducted emissions
CISPR11
Radiated emissions
CISPR11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/Flicker
emissions IEC 61000-3-3
Printed copies are uncontrolled - Check you are using the latest version
P01_0020 - USER MANUAL
Table 1 : Emission tests for all ME EQUIPMENT and ME SYSTEMS
Compliance
AKTIIA bracelet uses RF energy only for its internal function.
Groupe
1
Therefore, its RF emissions are very low and are not likely to
Class B
cause any interference in nearby electronic equipment.
Not Applicable
-
Not Applicable
Environment", i.e.
all establishments, including domes-
Electromagnetic environment - guidance
Version: 2
.
Page 11 of 14