C800 Analyser Operating Manual

3.2.3. Direct Connection to Analyser

To wire directly to an external power source, wire +24V into the +ve terminal and 0V to the -ve
terminal. Use 2-core screened cable, such as Belden 8760, through a gland in the bottom of the
analyser. Only connect the cable screen at the power source, not in the analyser.

3.3. Sensor Wiring

Connect a Model 300 conductivity sensor through a gland in the bottom of the analyser.
Connection
Conductivity Signal
Conductivity Signal
Pt100 Temperature Signal
Pt100 Temperature Reference
Pt100 Wire Compensation (Optional)
If you are using a 3-wire temperature compensator, connect the cable compensation wire to RTD
COMP and remove jumper J1.
902857-A - C800 Analyser Operating Manual.docx
Terminal
COND 1
COND 2
RTD SENSE
RTD RETURN
RTD COMP
Model 300 Wire Colour
Red
Black
Green
White
N/A
7