2. If the displayed value is not within acceptable tolerance (see chart below), adjustment is required.
VIII.
Adjustment
1. Turn the scale OFF.
2. Press and hold the UNITS key until the display shows CAL then release
3. Press the UNITS key again. The display will flash CAL followed by the required calibration weight (see chart).
4. Place the required weight on the center of the platform.
5. Once the weight is confirmed, the display will show "PASS" and return to normal weighing. Calibration complete!
IX. Troubleshooting
•
Problem: Scale will not turn on
•
Solution: Check the batteries
•
Problem: Displayed weight fluctuates randomly
•
Solution: Make sure your workspace is stable and free of any air currents or vibration
•
Problem: Displayed weight is inaccurate
Solution: Perform a calibration adjustment
•
X. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration adjustment. If the problem persists, the load cell may be
permanently damaged due to overloading.
2. UNST - Scale is unstable. Use in a more stable position.
3. HHHHH - Maximum capacity exceeded. Remove the extra weight to avoid damaging the load cell.
4. L - Batteries are low. Replace the batteries.
XI. Specifications
Max. Capacity
Readability
Power
Dimensions
Platform Dimensions
Warranty
Model
Acceptable
Tolerance (±)
CD-V2-100
CD-V2-500
CD-V2-100
100g / 3.527oz / 500ct / 1543.2gn
0.01g / 0.001oz / 0.05ct / 0.1gn
Calibration
Weight
0.03g
100g
0.2g
500g
500g / 17.635oz / 16.075ozt / 321.5dwt
0.1g / 0.005oz / 0.005ozt / 0.1dwt
2 x CR2032 Lithium Batteries
0.5 x 4.9 x 5.5"
4" diameter
10 Year Limited Warranty
Recommended
Accuracy Class
OIML M2
OIML M2
CD-V2-500