Amersham Pharmacia Biotech M-925 Manual de instruções - Página 5

Procurar online ou descarregar pdf Manual de instruções para Misturador Amersham Pharmacia Biotech M-925. Amersham Pharmacia Biotech M-925 8 páginas.

Technical specifications

Operating data
Max Flow rate
0.001 – 100 ml/min
Max Pressure
12 ml chamber 10 MPa at 0.1 – 100 ml/min
(100 bar, 1450 psi)
0.6, 2 and 5 ml chambers
25 MPa at 0.001 – 100 ml/min
(250 bar, 3625 psi)
pH range
1–13, 1–14 (<1 day exposure)
Viscosity
Max. 5 cP
Environment
+4 to +40 °C
20-95% relative humidity
84-106 kPa (840-1060 mbar)
atmospheric pressure
Physical data
Internal volume
4 interchangeable mixing
chambers: 0.6, 2, 5 and 12 ml
Mixing principle
1 static chamber and
1 dynamic chamber
Degree of
protection
IP 43
Wetted materials
Stirrer
PTFE (polytetraflouroethylene)
Mixing
chambers
PEEK
Chemical
resistance
The wetted parts are resistant to
organic solvents and salt buffers
commonly used in
chromatography of biomolecules,
except 100% ethylacetate, 100%
hexane and 100%
tetrahydrofuran (THF)
pH stability range
1–13, 1–14 (<1 day exposure)
Power requirement
12–40 V DC from the pump
Power consumption
1 W
Inlet and
outlet tubing
UNF 10-32 2B "Fingertights"
with capillary tubing 1/16"
outer diameter
Dimensions,
H x W x D
152 x 77 x 111 mm
Weight
1.2 kg
Mixer M-925
Instructions 56-8101-00 Edition AC
EMC Standards
This product meets the
requirement of the EMC Directive
89/336/EEC through the
harmonized standards
EN 50081-1 (emission) and
EN 50082-1 (immunity)
Note: The declaration of
conformity is valid for the
instrument when it is
• used in laboratory locations
• used in the same state as it
was delivered from Amersham
Pharmacia Biotech except
for alterations described in the
user manual
• connected to other CE labeled
Amersham Pharmacia Biotech
instruments or other
products as recommended.
5