Cognionics CGX Mobile-128 Manual - Página 17

Procurar online ou descarregar pdf Manual para Equipamento médico Cognionics CGX Mobile-128. Cognionics CGX Mobile-128 20 páginas.

Cognionics CGX Mobile-128 Manual
Software Reference
Channels Tab
The bottom half is a graphical map of sensor contact
quality. Red circles indicate poor contact, whereas light
green circles indicate acceptable contact. Dark green
circles indicate ideal contact.
The threshold for this contact quality is, by default:
Above 5000 kΩ
Red
Between 2500-5000 kΩ
Light Green
Below 2500 kΩ
Dark Green
This threshold can be changed with the Mask Hi-Z menu
on the display tab.
100% or greater of Hi-Z threshold
Red
Between 50%-100% of Hi-Z threshold
Light Green
Below 50% of Hi-Z threshold
Dark Green
17
The top half of the Channels
tab contains a table
displaying information about
each channel in the system
including its number, name,
contact impedance (Z (kΩ))
and sensor offset (Off (mV)).
The offset is calculated
relative to the reference
channel A1.