Amphenol Temposonics GP Manual do utilizador - Página 6
Procurar online ou descarregar pdf Manual do utilizador para Acessórios Amphenol Temposonics GP. Amphenol Temposonics GP 8 páginas. Linear position sensor analog and digital pulse outputs
Também para Amphenol Temposonics GP: Instrução breve (12 páginas)
Programming & Diagnostics Using a PDA - G-Series Digital Pulse Output Sensors Only.
Reading Sensor Parameters (See Figure 4)
Using the PDA configuration program it is possible to read the sensor
parameters listed in Appendix A on page 8. To view these parameters or sta-
tus variables, in the PDA screen select Select Items Here (just under the
Begin Prog button) and select one of the individual parameters on the scrolling
list. To view the parameter select the < Get button.
For a complete parameter listing, click the Get all Items button and then click
the < Get button and all the parameters and their values will stream across
the program display box. To scroll through the list, tap the up and down
arrow keys on the right of the screen. The PDA screen can be cleared of all
parameters by clicking the Clear List button.
Reading Magnet Position (See Figure 5)
The approximate magnet position can be displayed (in mm) to help with sen-
sor setup, or during troubleshooting of a system installation. The values dis-
played are for diagnostic purposes and are NOT representative of the resolution or
performance of the G-Series sensor. The PDA has a 3 second update time for
displaying magnet position.
1. To read magnet position, click the Start Reading button from the
main G-Series Configuration screen. The magnet position will appear on
the PDA screen.
2. To exit this mode, click the Stop Reading button.
Set (Programming) Mode (See Figure 6)
The Set Mode command is used to program the Digital Pulse output sensor
parameters.
1. From the main G-Series Configuration screen, select Set Mode. If the
PDA is properly aligned for IR communication, the current output
parameter values will be displayed. If not, position the PDA for IR
communication and select Get Mode Params.
To modify the output parameters simply select the pull-down menus and
select the desired parameter values. Use the down arrow on the right to
view all parameter values available for the following items:
• Output Mode – Start/Stop or PWM
• Interrogation Type – External or internal interrogation for PWM mode
• Number of recirculations for PWM mode (1 to 15)
2. Once finished with parameter selection, save settings to the sensor by
selecting Send to the right of the mode parameter menu. The
"Status" line at the bottom of the screen will display "Command
Complete" when changes have been successfully applied to the sensor.
3. To exit the Set Mode, click the Return to Get Mode button to return to the
main menu. Power to the sensor must be cycled for mode changes to
take effect. If changes are not saved prior to removing sensor power, the
parameters will revert to their last saved values.
Figure 4 - Selecting Sensor Parameters
Display box
Parameter list
IR communication status
Figure 5 - Reading Magnet Position
Figure 6 - Using Set Mode
Programmable output
parameters
Return to main menu
Get current parameter values
6
Begin IR communications
Magnet position
Output parameters
Read parameter values
Magnet position reading
Return to main menu
Zero magnet position output