BQ610-B Online Controller Operation manual

Calibration

PC tool name: SmartPC(DO).exe.
Note: If need any help, please contact Customer Service.
Calibration Solution Preparation

Labware and materials required

Analytical balance,
1L volumetric flask, 1L brown flask
Pipette
NDSA (1, 5-Naphthalene disulfonic acid disodium salt dihydrate), CAS: 76758-30-0
Stock Solution Preparation
Wise 0.1g NDSA, dissolve it in deionized water, transfer it to volumetric flask, constant volume to
1L, and get 100ppm NDSA stock solution.
Different concentrations of standard solution preparation
Dilute 100ppm NDSA stock solution according to the required concentration. For example, take
50mL 100ppm NDSA stock solution and volume it with deionized water to 1L, then get 5ppm
NDSA standard solution.
Note:
Please operate carefully in the process of configuring standard liquid, and wear
gloves;
The test should be carried out in a brown light-proof bottle;
In the calibration process, please use deionized water to clean the sensor and dry
it to avoid contaminating the standard solution.
50mL beaker, 100mL measuring cylinder, glass rod,
Lab spoon,