Amprobe VPC-20 Manual de instruções

Procurar online ou descarregar pdf Manual de instruções para Instrumentos de medição Amprobe VPC-20. Amprobe VPC-20 2 páginas. Digital voltage and continuity tester

Digital Voltage and Continuity Tester
Instruction Manual
Safety Information
• Never use the tester if it or test leads appear damaged.
• Never apply more than 600V DC or 600V AC rms. between any terminal and
earth ground
• When using the probes, keep your fingers behind the finger guard on
the probes.
• Turn off power to the circuit under test before cutting, desoldering, or
breaking the circuit. Small amounts of current can be dangerous.
• Use caution when working with voltage above 60V dc or 30V ac rms. Such
voltages pose a shock hazard.
Automatic Operation
The tester automatically turns on when you short the two test probes or place the
probes across a complete circuit.When turned on, the tester powers up in continuity
mode. If a dc or ac voltage greater than about 4.0V is present across the inputs, the
tester switches to dc or ac voltage mode.The tester automatically shuts down after
approx. 4 minutes of inactivity.
Continuity testing
Turn off circuit power before testing. Beeper indicates shorts lasting 1 ms or longer.
Measuring Voltage
Refer to Automatic Selection
DC Voltage (Volts dc>4.0V)
600V dc maximum
SHORT 10K
Model VPC-20
AC Voltage (Volts dc>4.0V)
600V ac maximum
50Hz to 400Hz
OPEN
12V DC
Specifications:
Function
Range
V AC(50Hz ~400Hz)
600V
V DC
600V
Continuity test: Audible threshold 0~10K
Overload protection: AC/DC 600V
Diminsions/wt.:215x96.3x37.5mm/312g
Display: Large 1999 count LCD display with function and high voltage warning indication.
Maximum Voltage Between any Terminal and Earth Ground: 600V dc; 600V ac rms,
Over-voltage Category III 600V
Safety: Complies with EN61010-1:1993 for use in over-voltage CAT III environments
Pollution Degree: 2
For indoor use only
EMC Regulations: EN50081-1, EN50082-1
Polarity: Automatic, (-)negative polarity indication.
Over-range: "1" mark indication
Low battery indication: The "BAT" is displayed when the battery voltage drops below
the operating level
Measurement rate: 2 times per second, nominal
Continuity test: Audible threshold 0-10KW, Continuity beeper 2KHz
Input Impedance: 400K
Temperature: 5 °C(41°F) to 40°C (104°F)
Altitude: Operating: 2000m; Storage: 10,000m
Relative Humidity: 80% for Temp up to 31C decreasing linearly to 50 % RH at 40C
Pollution Degree: 2
For indoor use only
Battery Type: AA (2) NEDA 15F or IEC R6
240V AC
Accuracy(%rdg.+digits)
+(2.9%+8)
+(1.5%+2)
Ω,Continuity beeper 2KHz.