Amprobe WT-30 Manual do utilizador - Página 11

Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição Amprobe WT-30. Amprobe WT-30 17 páginas. Salinity meter

Amprobe WT-30 Manual do utilizador
Salinity calibration
1. Dip the probe into demineralized or distilled
water for about 30 minutes to rinse the probe.
2. Select the NaCL standard for calibration.
3. Pour enough solution into two separate clean
containers.
4. Power on the meter.
5. Rinse the probe into one of above containers.
Gently stir the probe.
6. Dip the rinsed probe into the second container.
Tap probe on the bottom of container to remove
air bubbles. Wait about 15 mins to make the probe
stabilize to the solution temperature.
7. Press "CAL" more than 2 seconds to begin
calibration. The salinity value of solution will blink
on LCD.
8. Press "p" and "q" to change the value in order
to match the value to the standard solution. You
can adjust the salinity reading for +30%. However,
if the measured value and standard value differs
by more than 30%, it is suggested to clean probe
or replace meter.
9. When the "CAL" stop flashing, you can
press"SET" less than 1 second to confirm and
return to salinity measurement mode. If "CAL"
always blinks, check if the calibration solution is
stable enough and whether the step 8 input value
is correct or not.
10. Repeat 1~9 for other ranges if needed.
11. To exit salinity calibration mode without
confirming, press "SET" in step 9 for more than 2
seconds.
6