TESTING
The Bio Cart HEPA-UVC 13 has mechanical, electronic, camera and display functions to test. If any of
these tests fail, contact Air-Care Immediately. 800-322-9919 or [email protected]. The Bio Cart HEPA-
UVC 13 may require the ladder to be removed to perform some of the tests.
Turn off all switches in the Bio Cart HEPA-UVC 13 before connecting to AC power.
1. Test the brakes on the 2 FRONT (Image 4) wheels with brakes to
ensure they will secure the cart in working position.
2. Test the swivel locks on the 2 BACK caster wheels . Lock the
swivel and engage a wheel brake at the same time.
3. 2 switches on the UVC lamps power boxes are
below the Pre-filter . (Image 5)
4. The "IN USE" light Switch on the door is on when
the Bio Cart battery switch is on whether the unit is
plugged in or unplugged.
5. Inspect the power plug for damage (Image 6)
6. Pull the power cord out several feet, give it some slack, and it
should lock in place.
7. Pull the cord tight again and allow it to be pulled in and rolled up
into the cart.
Testing Continued Page 8
PAGE 7
(Image 4)
(Image 5)
(Image 6)