5.2. Amptek SiPIN Specifications
General
Detector Type
SiPIN
Detector Size (mm
2
) 6
2
Collimated Area mm
4.4
Thickness
500μm
Collimator Type
Internal Multilayer
Preamplifier Type
JFET with Amptek custom reset
through HV
Energy resolution @ 5.9 keV (55Fe)
6 mm
2
:139 to 159 eV FWHM at T
13 mm
2
:150 to 205 eV FWHM at T
25 mm
2
:190 to 225 eV FWHM at T
Other Performance
Peak to background
16,000:1 for 6 mm
(typical)
15,000:1 for 13 mm
> 12,000:1 for 25 mm
Maximum input
> 50 kcps
count rate
Throughput
Determined
count rate stability
processor & its settings
Window Options
Be
1 mil (25
0.5 mil (12.5
1 μm Parylene on each
Signal Output
Sensitivity
1 mV/keV
Gain stability
<20 ppm/°C
Polarity
Negative signal
Reset range
-5 to +5 V (typical)
Cooling
Cooling performance ΔT > 85°C
Cooler type
Two stage thermoelectric
Temp monitor
Diode
Power
HV Bias
+ 180V (typical)
Max cooling power
3.5 V / 0.45A
Total power
< 2 W (full cooling)
X-123 User Manual
13
25
11.1
21.5
=32
s
pk
=32
s
pk
=32
s
pk
2
2
2
by
signal
m)
m)
Other
Operating range
Storage & shipping
RoHS
Lifetime
Warranty Period
Preamplifier Options
Amptek's Si-PINs are available with the XR-100
preamplifier, with the PA210 or PA230 OEM
preamplifiers, or as part of the X-123 and X-55.
Signal Processing Options
Amptek's Si-PINs may be used with Amptek's DP5 or
DP5-X, or the PX5, or as part of the X-123 or X-55.
Specification Notes
Performance listed here is measured at full cooling
(<220K).
The "peak to background ratio" is the ratio of counts
in the Mn Kα peak channel to the counts at 1 keV.
"Gain stability" refers to the change in sensitivity
when the preamp temperature changes but the
detector temperature is stable.
Operating Notes
Detector must be operated with an appropriate heat
sink.
The SiPIN is suitable for vacuum applications.
Amptek provides several system configurations.
9
-35
o
C to +80°C Performance
degrades at elevated detector
temperatures.
-40
o
C to +85°C, 10% to
90% RH noncondensing
Compliant
Typical 5 to 10 years, depending
on use
1 year