EN
User manual - English
4. Operating the Mach LED 150 light
ON/OFF switch of the light
(to switch off, hold for one sec-
ond) (1)
1
2
Adjust the brightness in six lev-
els between 50 and 100%
(2 and 3) (lowest leve l= endos-
copy light)
4
Display of the set brightness (4),
the uppermost LED will flash in
3
Standby
If only the top LED lights up, the
light is in endoscopy mode
Focus the light field by turning
the handle (5)
5
Position the lamp body using
the handle (5) or the handle
bar (6)
6
5
59321001
5. Operating the Mach LED 150MC light
Switching the light on/off
(to turn off the light, press and
hold the (ON/OFF) button for one
second).
Adjusting the brightness
Switching on the endoscopy
mode
Adjusting the colour temperature

6. Cleaning and disinfecting

Cleaning and disinfecting work must only be done by trained
staff. The respective requirements must be observed for all
cleaning and disinfection work (details can be found in the man-
ual).
Housing/protective screen
The housing and the protective screen of the lamp body can be
cleaned and disinfected with many common/commercially available
materials. Do not use cleaning agents or disinfectants contain-
ing active substances based on biguanides, phenols, chlorides
or percompounds.
Furthermore, only cleaning agents approved for polycarbonate (PC)
may be used to clean the protective screen. To protect against me-
chanical damage, always use a damp cloth (never a dry one) to wipe
the protective screen and after cleaning, wipe with an anti-static
agent (lint-free cloth).
Sterilisable handle
The handle must be cleaned/disinfected before each use. It can be
steam sterilised (max. 200 sterilisation cycles for max. 5 minutes at
a max. temperature of 134°C); details can be found in the manual.
Before installing the handle, check it for visible damage, fouling
and the specified manufacturing date. Do not use damaged or
dirty handles or handles that are more than two years old.
8 / 93
7. Faults
In the event of unusual examination conditions, the lamp must not be
used, as safe operation cannot be guaranteed. For troubleshooting,
disconnect the lamp from the mains for about 30 seconds. In the
event of continuous faults, a suitably trained service technician must
be contacted.
8. Information on electromagnetic
compatibility
Medical electrical equipment is subject to special precautionary
measures with regard to electromagnetic compatibility (EMC). It may
only be installed and put into service in accordance with the EMC
instructions in the accompanying documents. The Mach LED 150
and Mach 150MC operating light are intended for use in professional
healthcare facilities.
The emission characteristics of this device
permit its use in industrial and hospital envi-
ronments (CISPR 11, Class A). When used in
a residential environment (which usually re-
quires Class B according to CISPR 11), this
device may not provide adequate protection of
radio services. The user may need to take re-
medial action such as relocating or reorientat-
ing the device.
Portable and mobile RF communications
equipment can affect medical electrical equip-
ment and must not be used within 30 cm of the
light, including the cable.
The light is suitable for use in conjunction with
an RF surgical device. A minimum distance of
50 cm must be maintained between the surgi-
cal light, including its suspension system, and
the RF electrode cables.
The use of this equipment immediately next to
other equipment or with other equipment in
stacked form should be avoided since this may
result in faulty operation. Should use in the
aforementioned manner nevertheless be nec-
essary, this device and the other equipment
should be kept under observation to ensure
that they are working properly.
Mach LED 150
V6