and the heating is operated by the position control.When self-tuning finished, the
instrument returns to the normal control state. If enters the setting state in the self-tuning
process, it will exit from the self-tuning program.
3)In the running or stop state, shortly press the "OK"key to enter the temperature setting
state, the upper row displays "SP", the lower row displays the setting temperature, then
press the "OK"key again, the upper row displays the "ST" and the lower row displays the
setting time.
4)Long press "OK"key for 5 seconds to enter the setting state of other parameters. The
upper row will display the prompt and the lower row will display the parameter values
(see the attached table for details of each parameter). After setting the parameter, long
press "OK"key for 5 seconds to exit the setting state. In each setting state, you can
quickly modify each parameter by pressing the  key,  key and the  key. After
modification, must press "OK" key to confirm, otherwise it is invalid.
Parameter specifications:
Symbol
SP
ST
Long press the OK key
Lck
SC
Lck is××××(the password can be obtained from the manufacturer)
P
I
d
CT
COOL
AL
PT
Drawell International Technology Limited
Web : www.drawell.com.cn
Value
SPL~SPH
0~9999 min
0~9999
-20.0~+20.0
0.1~20.0
1~400
0~100
60~600
SPL~50.0
1.0~5.0
1~60 min
Tel: +86-023-63268643
Description
Temp. setting value
Time setting value
LOCK
Correction for measured
value
Output proportional
Integral Time
Derivative time
Compressor start delay
Upper temperature limit for
compressor normally open
Upper deviation alarm
Print interval
Factory set
value
random
random
——
——
——
——
——
——
0℃
——
——