BD 9003TIG03-G Manual de referência rápida - Página 7
Procurar online ou descarregar pdf Manual de referência rápida para Equipamento médico BD 9003TIG03-G. BD 9003TIG03-G 18 páginas. Volumetric pump

VOLUME
VOLUME INFUSED
VTBI
SET VTBI WITH
OK
ON HOLD
RATE
!
SET RATE WITH
VOLUME
0000CF04424.indd 7
6.
Clear Volume Infused, if required, by selecting VOLUME
softkey and then press CLEAR softkey.
374
ml
QUIT
CLEAR
7.
Enter Volume To Be Infused (VTBI), if required, by
VTBI
PRIMARY
selecting VTBI softkey on main display.
(OFF)
0
•
.0ml
•
QUIT
BAGS
8.
Enter or adjust the rate using the f keys, if
PRIMARY
required.
0
.0ml/h
9.
Connect the fully primed infusion set to the patient's
vascular access device.
10. Press b key to start the infusion. INFUSING will be
VTBI
displayed.
Set VTBI using the BAGS option and/or the f
keys. Press OK to confirm.
Set the END RATE using the f keys. Press OK
to confirm.
To stop the infusion immediately perform one of
the following actions:
• press the
h
key, recommended action
• close the roller clamp and then open the door
0000CF04424 Iss. 1 7/18