BD LSR II Manual do utilizador - Página 4

Procurar online ou descarregar pdf Manual do utilizador para Instrumentos de medição BD LSR II. BD LSR II 8 páginas. Flow cytometry analyzer
Também para BD LSR II: Manual do utilizador (20 páginas)

BD LSR	II Manual do utilizador
c. Repeat for each parameter you're not using
d. Do not delete Parameters that may be of interest.
e. If you add back a deleted parameter to the experiment it will come
in with different voltages, not specified PMT settings by facility.
This may cause problems for resolution. Check cytometer
settings for the 197 bead tube to see correct PMT voltage to
assign when adding back a parameter.
4. Keep forward and side scatter height and width selected
5. The data may be viewed compensated or not, but the uncompensated
data is written to the database along with the matrix so that other
compensations may be used later in FlowJo.
Note: For each tube collected in Diva software, uncompensated data is
recorded and the compensation matrix is applied as a transform, it does not
affect your data.
All samples must be filtered for sorters. The analysis instruments have nylon
mesh squares on the desktop if you need them
Compensation
• Corrects for the emission of one fluor into the detector used to measure another.
• Flow Cytometer use filter sets that are optimized for very specific wavelengths of light.
• Dye/Fluorochrome emissions often span from one filter into another.
• COMPENSATION IS VERY IMPORTANT - without it, our data cannot be interpreted properly.
• To calculate compensation, every experiment needs an unstained control as well as each
fluorochrome singularly.
• Antibody capture beads can be substituted for cell controls.
1. Create Compensation Controls. Select Experiment à
Compensation Setup à Create Compensation Controls.
2. In the Create Compensation Controls
window, ensure that all of your colors are
labeled "Generic". Then select "OK". Do
not compare negative cells to positive
beads or vice versa.
4