Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, request service from your dealer
or from the A&D service group. The A&D service group will provide
technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the monitor inspected
every 2 years to ensure proper functioning and accuracy. This
maintenance recommendation is for European customers only, to comply
with regulatory requirements. Routine inspection and maintenance is not
needed for non-European customers. Please contact either your
authorized dealer or A&D for maintenance.
Type
Measurement method
Measurement range
Measurement accuracy
Power supply
Upper arm circumference 22 - 32 cm (8.7" - 12.6")
Classification
Clinical test
EMC
Wireless communication
Operating condition
Storage condition
Dimensions
Weight
Note: Specifications are subject to change for improvement without prior notice.

Maintenance

Technical Data

UA-767PBT-C
Oscillometric measurement
Pressure: 20 - 280 mmHg
Pulse:
40 - 200 beats / minute
Pressure: ±3 mmHg or 2%, whichever is greater
Pulse:
±5%
4 x 1.5V batteries (R6P or AA)
Type BF
According to ANSI / AAMI SP-10 1987
IEC 60601-1-2: 2001
WML-40AH (MITSUMI Electronics Co. Ltd.)
Bluetooth Ver.2.1 Class 1 HDP
Continua certified
+10°C to +40°C / 30%RH to 85 %RH
-10°C to +60°C / 30%RH to 85 %RH
Approx. 147 [W] x 64 [H] x 110 [D] mm
Approx. 300 g, excepting batteries
13