5.5 Re‐Calibrating a Site
1. When in the Site/Sensor Details
screen, using the left‐right keys,
scroll to Calibrate and select.
2. Follow instruction to calibrate.
6 Maintenance
CAUTION: Always disconnect power from the instrument before attempting any cleaning operations.
Important: Under no circumstances should the instrument, display or the accessories be cleaned with solvents
such as white spirit, acetone, etc.
Fluorimeter
‐
Clean the enclosure, sample cell compartments and all accessories with a soft damp cloth. A mild soap
solution can also be used. Do not get excess water in the sample cell compartments. Do not insert a brush
or sharp object into sample cell compartment to avoid damaging the mechanical components. Dry the
cleaned parts carefully with a soft cotton cloth.
Display
‐
Take care not to scratch the display. Do not touch the screen with ballpoint pens, pencils or similar pointed
objects.
‐
Clean the display with a soft, lint‐free and oil‐free cotton cloth.
‐
Diluted window cleaner liquid can also be used.
Sample Chamber
‐
Splashes or spills on and in the instrument should be cleaned up immediately. Remove any liquid inside the
sample chamber by using a non‐lint swab. Do not tip the instrument to empty liquid from the sample
chamber.
Sensors
‐
Never reuse sensors! Once the sensor bag is exposed to air, it should be used immediately (within 30
minutes) as exposure to humidity, air and/or heat can affect its effectiveness. Dispose used sensor and all
other used consumables such as sample tubes, syringes, plastic pipettes after each analysis.
Pipettes
‐
The 100µL automatic pipette can be reused as it uses disposable pipette tips. Pipette tips should be
disposed of immediately after use.