ATI Technologies Q46P/R Manual de Comunicações - Página 15

Procurar online ou descarregar pdf Manual de Comunicações para Instrumentos de medição ATI Technologies Q46P/R. ATI Technologies Q46P/R 18 páginas.

ATI Technologies Q46P/R Manual de Comunicações
ATI
Q46 Ethernet/IP Communications Manual
1.9

Ethernet/IP Data Files

The data organization shown in section 1.6 corresponds to the same mapping as
all the other communication interfaces on the Q46. All Q46 instruments provide
20 bytes of input data to the master, and the meaning of the bytes changes
slightly depending on which Q46 instrument is used.
Byte
1 to 4
5 to 8
9 to 12
13 to 16
17
18
19
20
Data Type
Sensor
long(32)
Main
long(32)
Main
long(32)
Aux1
long(32)
Aux2
Char(8)
NA
Char(8)
NA
Char(8)
NA
Char(8)
NA

Figure 9 - Q46 Ethernet/IP Register Data Format

Byte
Bitfield
17
0 (lsb)
17
1
17
2
17
3
17
4
17
5
17
6
17
7
18
0
18
1
18
2
18
3
18
4
18
5
18
6
18
7
19
0
19
1
19
2
19
3
19
4
19
5
19
6
19
7
Figure 10 – Base Status Register Bitfield Flags
Description
Measured Main
Measured
Temperature
Measured Aux1
Measured Aux2
System Status 1
System Status 2
Alarm Status
Instrument ID
Description
1
Raw Hi
Raw Lo
1
Main Hi
Main Lo
Temp Hi
Temp Lo
Aux1 Hi
2
Aux1 Lo
NU
NU
LCD Controller Fail
Cal Main Fail
PID Controller Fail
Cal TC Fail
TC Error
Acknowledge Fail (global)
Alarm 1, Relay A
Alarm 2, Relay B
Alarm 3, Relay C
Alarm 4, Relay D (optional)
Alarm 5, Relay E (optional)
Alarm 6, Relay F (optional)
NU
NU
15
Data Format
See Figure 11
25.00° C = 2500
See Figure 11
See Figure 11
(binary) 00000000
(binary) 00000000
(binary) 00000000
See Figure 11