ATI Technologies B12-69 Manual O & M - Página 8
Procurar online ou descarregar pdf Manual O & M para Transmissor ATI Technologies B12-69. ATI Technologies B12-69 20 páginas. 2-wire wet oxidant gas transmitter
Model B12-69 Wet Oxidant Gas Transmitter
Explosion-proof transmitter enclosures should be supported by the conduit system and the
installation should include proper conduit seals as required by local electrical code.
Figure 4 - Expolsion-Proof Transmitter Enclosure Dimensions (ATI-0226)
The back of the insertion sensor is a 1¼" MNPT pipe thread. The sensor is normally threaded
into a blind flange or pipe saddle which is used to mount to the vent. The sensor should be installed
perpendicular to the air flow and should be inserted far enough into the vent so that the sensor tip is at
least ½" into the stack. The 1¼" adapter for sensor mounting must be supplied by the user.
The sensor in a flow configuration is used in conjunction with the low volume flowcell (see Figure
2) to provide for sample to be delivered to the sensor using a pump or the sample pressure from the vent
stack. The sensor simply threads into the low volume flowcell and ¼" hose barb fittings are provided for
inlet and outlet connections.
Figure 5 - Wet Oxidant Gas Sensor Dimensions (ATI-0276)
O & M Manual
Rev-G, 7/15
- 8 -