The meter for single-patient use and the lancing devices
may be cleaned 5-6 times per day (Test 5-6 times every day),
so the meter will be cleaned and disinfect about 11000 cycles
that include cleanings plus disinfections (1 cycle = 1 cleaning
+ 1 disinfection) during its 5 years life cycle.
1. Before the test, the user should clean and wash hands. If you
are a caregiver, (not the patient), the please use disposable
gloves.
2. After test, take one wipe to clean the meter's housing, buttons,
lens, mark, usb cover and support block, including its front and
back for 10 seconds for cleaning until there is no soil on the
surface. Then disinfect the meter's housing, buttons, lens, mark,
usb cover and support block use another wipe for 2 min, keep
the surface wet for 2 min.
3. Take off the glove; discard carefully according to your health-
care provider's instruction.
Note:
1. Each disinfection step requires a pre-cleaning step.
2. If the meter is being operated by a second person who is pro-
viding testing assistance to the user, the meter and lancing
device shall be cleaned and disinfected first.
3. The users should wash hands thoroughly with soap and water
after handling the meter, lancing device, or test strips.
1. Model: ADF-B27 / ADF-B27A*
2. Machine size: 90mm (L)×54mm (W)×13mm(H)
3. Weight: 73g (excluding batteries)
4. Measuring method: Amperometric technology using glucose
oxidase
5. Result range: 20 mg/dL ~600 mg/dL(1.1 mmol/L ~33.3mmol/L)
6. Glucose Units: Either mg/dL or mmol/L*(ADF-B27A)
7. Memory : 500 blood glucose tests
8. Display: LCD display with backlight
9. Power source: DC 3V (AAA×2 alkaline batteries)
10. Battery Life: Approx. 500 normal tests
11. Storage Condition: Blood glucose Monitoring system (meter
and test strips): 39.2
12. Storage Condition (meter): -4
13. Operating conditions: 50
14. Blood Source: Fresh capillary whole blood
15. Blood Volume: Minimum 0.7 micro liter.
16. Use life: five years
Limitations of use:
※ The blood Monitoring System is not intended for use on neo-
nates.
※ The blood Monitoring System is not intended for use on
artery blood, neonates' serum and plasma.
※ The Blood Glucose Meter should be used with " Andesfit "
test strip.(ADF-B27GTS)
※ The system can be used up to an altitude of 10744 feet.
※ The following substances at levels greater than normal or
therapeutic levels may cause significant interference (affect
the result by greater than 10%) resulting in an inaccurate
104 , Humidity<80%RH.
131 , Humidity<80%RH
104 , Humidity:25%~80%RH