Andon KD-5904 Manual de operação - Página 13

Procurar online ou descarregar pdf Manual de operação para Monitor de tensão arterial Andon KD-5904. Andon KD-5904 19 páginas.

DN:KD-5904-SMSY01 V3.0
Note: It is not intended to provide a basis of any type of rush toward emergency
conditions/diagnosis based on the color scheme and that the color scheme is meant only
to discriminate between the different levels of blood pressure.

13. TROUBLESHOOTING

PROBLEM
LCD Display shows battery
symbol
LCD Display shows "EE 0"
LCD Display shows "EE 1"
LCD Display shows "EE 2"
LCD Display shows "EE 3"
LCD Display shows "EE 4"
LCD Display shows "EE 5"
LCD Display shows "EE 6"
LCD Display shows "EE 7"
LCD Display shows "EE 8"
LCD Display shows "EE 9"
LCD Display shows "EE A"
LCD Display shows "EE b"
POSSIBLE CAUSE
Batteries are low
Pressure system unstable
before measurement.
Systolic pressure detect fail
Diastolic pressure detect fail
Pneumatic system blocked or
cuff is to tight during inflation
Pneumatic system leakage or
cuff is to loose during inflation
Inflation above 300mmHg
Over 3 minutes while pressure
above 15 mmHg
EEPROM access error
Device parameter check error
MCU self-verity error
Span error
EEPROM backup error
13/19
Operation Guide
SOLUTION
Install new batteries
Re-test taking care to not
move your arm or the blood
pressure monitor
Measurement again
Measurement again
Tighten cuff correctly and
refer to "CUFF Connecting"
Tighten cuff correctly and
refer to "CUFF Connecting"
Re-test taking care to not
move your arm or the blood
pressure monitor
Press the button "START"
Please contact the factory
Please contact the factory
Please contact the factory
Please contact the factory
Please contact the factory
JS-KD5904-03-V2.0