Andor Technology Marana CXP Manual de início rápido

Procurar online ou descarregar pdf Manual de início rápido para Instrumentos de medição Andor Technology Marana CXP. Andor Technology Marana CXP 2 páginas.

Andor Technology Marana CXP Manual de início rápido
Marana Quick Start Guide - USB3 and CXP
1
Prepare your Facility
Handle the camera and accessories with
care.
Observe anti-static precautions when
installing the controller card and camera.
Ensure 100 mm clearance around camera
and power supply vents (Do not block vents
when operating on benchtop).
Operation of the camera close to intense
pulsed sources (e.g. plasma and X-ray
sources) may compromise performance.
Keep dry, <70% humidity non-condensing.
Use correct power supply and cables.
Store between -10°C and 50°C.
Operate between 0°C and 30°C.
Cooling options information:
Min. temp with air cooling: -25°C.
Min. temp with water/liquid cooling: -45°C.
If using water/liquid cooling the coolant
system must be connected and coolant
temperature set above the dew point before
the camera is switched on.
Refer to the user manual for further information
2
Unpack the Camera
Camera
Essentials
USB 3.0
CoaXPress
PCIe Card
Card
(with ext. trigger
input)*
Marana
Back-illuminated
sCMOS Camera
(model as ordered)
Documentation
Software
scan or visit:
(copy supplied)
andor.oxinst.com/
welcome
3
Install Software (Solis or SDK3)
You must have administrator access on your PC to perform this installation.
These instructions cover installation of both Andor Solis and Andor SDK if ordered. A copy of the software is provided
within the camera box. Additionally, the latest software files are available to download on request, please sign in or
create an account at
andor.oxinst.com/downloads
Windows: Switch on the PC, run the setup exe from the supplied copy or the downloaded file: e.g. AndorSDK3Setup-
<version>.exe.
1. Confirm the version of software (minimum is v3.14.x).
2. Follow the on screen prompts.
3. Select the installation directory when prompted.
4. If prompted for camera type select Marana.
5. On the final window click on "Install". Continue to follow the on screen prompts.
6. To complete the installation, when prompted select "Yes, restart the computer now" and click on the finish button.
Linux:
For detailed instructions on how to install the SDK3 and CoaXPress drivers on Linux please refer to Section 1.5.2 in
the Andor Software Development Kit Manual, available at
Cables
Anti-Static
USB 3.0 Cable
CoaXPress
Strap
Cable*
Coolant Pipes
Trigger Cable
Trigger Cable*
(hose inserts)
1 x BNC to D-type
1 x BNC to SMB
*CoaXPress models only
andor.oxinst.com/downloads
Country
Specific Power
Cord
Power Supply