Ankom XT15I Procedimento de serviço
Procurar online ou descarregar pdf Procedimento de serviço para Equipamento de laboratório Ankom XT15I. Ankom XT15I 5 páginas. Extractor sight glass assembly
1.
Remove the back cabinet.
a. Remove the eight screws (four on each side of the
instrument) that secure the back cabinet onto the
instrument using a 1/8" Allen wrench. Set aside
the screws for later use.
b. Lift up, then pull back so the handles on each side
slide out of the notched groove.
2.
Remove the front plate by unscrewing the four
attachment screws with a 1/8inch Allen wrench. Set
the screws aside.
ANKOM Technology • 2052 O'Neil Road •
Sight Glass Assembly (X85)
XT15/XT15I
https://www.ankom.com/contact/technical-services
Screws
Notched groove
Front
plate
screws
• Phone: 315-986-8090
#021
Service Procedure
Revised: 11/1/22
pg. 1 of 5