Safety:

- The centrifuge should be placed on a stable, flat surface.
- Before using the centrifuge, check the rotor. Do not operate the centrifuge if
the rotor is loose, or does not rotate freely.
- Rotors and accessories that show evidence of damage or corrosion or are to
be taken out of service.
- The centrifuge must be taken out of service if safety related damages are
found in the rotor chamber.
- Do not open the lid until the rotor has come to a complete stop.
- For use only on smooth level surfaces, or approved centrifuge stands.
- Do not move the centrifuge while it is plugged in or running.
- Do not operate the centrifuge in a moving automobile.
Lid Safety Switch:
The lid is secured to the top of the cabinet by a latching knob and pawl system.
When the knob is rotated clockwise, the pawl grips the underside of the cabinet
opening and prevents the lid from opening. A mechanical stop positions the
pawl and prevents it from rotating completely. When rotated to the stop position,
the pawl makes contact with a switch that is mounted underneath the cabinet
top. This "lid safety switch" prevents the centrifuge from operating while the lid
is open.
Over Current Protection:
The centrifuge is provides protection from overcurrents with a fused power cord.
In the event of an overcurrent condition, the fuse will blow - cutting power to the
centrifuge, while protecting the internal electronics.
Page 10