Drucker Diagnostics Horizon 12 Manual de serviço - Página 5

Procurar online ou descarregar pdf Manual de serviço para Equipamento de laboratório Drucker Diagnostics Horizon 12. Drucker Diagnostics Horizon 12 12 páginas.

Drucker Diagnostics Horizon 12 Manual de serviço
LOW SPEED ERROR
2 ERROR BUZZES + 2 REPEATING BEEPS
OVERCURRENT ERROR
3 ERROR BUZZES + 3 REPEATING BEEPS
NO RPM ERROR
6 ERROR BUZZES + 4 REPEATING BEEPS
EEPROM ERROR
2 ERROR BUZZES + 6 REPEATING BEEPS
ACCELEROMETER ERROR
2 ERROR BUZZES + 7 REPEATING BEEPS
MOTOR DRIVE ERROR
2 ERROR BUZZES + 9 REPEATING BEEPS

7. SERVICE INSTRUCTIONS

7.1. Cleaning
The cabinet, rotor and accessories can be cleaned using soap and water, isopropyl
alcohol, or a 10% (5500 PPM) bleach solution
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in
water or other cleaning solutions as this will cause damage and void the warranty.
Under no circumstances should any of the following be used: TBQ Germicidal
Products, Fully/Partially Halogenated Hydrocarbons, Ketones or Esters.
The use of any chemicals not prescribed by the manufacturer may cause damage
to the rotor and tube carriers and shall not be used.
7.2. Removing the Rotor
Use a 1/8" hex key to loosen the center rotor screw (turn counter-clockwise).
Lift the rotor straight up and out of the rotor chamber.
To install the rotor, reverse steps A and B above. Take care to align the shaft cross-
pin with the rotor hub. Tighten the rotor screw to 3.5 Nm.
7.3. Maintaining the Rotor
Keep the rotor clean, any corrosive materials must not be allowed contact with the
rotor and should be cleaned immediately.
The rotor should be checked periodically for signs of wear.
Remove the rotor from service if any of the following are found: cracks, deep
scratches, corrosion or discoloring.
7.4. Rotor Screw
If the rotor screw needs to be tightened, use a 1/8" hex key and tighten to 3.5 Nm.
7.5. Speed Calibration
Check the centrifuge speed periodically, every two years is recommended.
Important: When verifying rotor speed, make certain that all carriers are installed
in the rotor.
No calibration adjustment of speed can be made, only a verification of rotor speed.
DRUCKER DIAGNOSTICS
HORIZON 12 SERVICE MANUAL
Rotor Speed is under set
speed for too long.
Motor or PCB failure
Sensor not connected to PCB
Sensor failure
Locked rotor
PCB failure
PCB failure
PCB failure
SM015
REV A
Check the lid and guard bowl gaskets. Voids in the rotor chamber
gaskets change the airflow, resulting in increased stress on the motor
drive
Service Required
Check connections
Replace motor
Service Required
Replace PCB
Replace PCB
Replace PCB
5