Target Diagram
FOCAL ZONE
Description and Reason For Testing
The focal zone is the region surrounding the focal point in which the intensity and the
lateral resolution is the greatest. Clinically, structures examined within the focal zone will
provide the best diagnostic information obtainable. The focal zone can be affected by
changes in the pulsing/receiving section of the imaging system or damage to the
transducer.
Test Procedure
1. Place the phantom on a clean, flat surface with the internal scanning well positioned
for use.
2. Fill the scanning well slowly with water to avoid introducing air bubbles.
3. Insert the transducer into the scanning well.
4. Adjust the instrument settings (TGC, output, etc.) to establish baseline values for
"normal" liver scanning. If the bottom of the phantom is seen, adjust the gain settings
until the image fades and goes entirely black at the bottom of the display. Record these