Absolute Process Instruments Cecomp DPG1000DR Series Instruções
Procurar online ou descarregar pdf Instruções para Transmissor Absolute Process Instruments Cecomp DPG1000DR Series. Absolute Process Instruments Cecomp DPG1000DR Series 2 páginas. 4-wire digital pressure transmitters
Cecomp
4-Wire Digital Pressure Transmitters
®
Ranges and Resolution
See table below for popular ranges. Consult factory for special
engineering units. Resolution is fixed as indicated. See our
F16DR series for ranges greater than 2000 or if more display
resolution is required.
† -HA option is for output only and not supported by display
‡ -HA option not available
PSI
inH
O
Res
Res
2
3PSIG
85INH2OG
‡
.01
‡
.1
5PSIG
.01 140INH2OG
.1
‡
‡
15PSIA
.01 400INH2OA
1
15PSIVAC
‡
.01 400INH2OVAC
‡
1
±15PSIG
.1 ±400INH2OG
‡
‡
1
15PSIG
.01 400INH2OG
1
30PSIA
.1
850INH2OA
1 ±1000CMH2OG
†
30PSIG
†
.1
850INH2OG
1
60PSIG
ftH
O
.1
Res
2
100PSIA
.1
7FTH2O
.01 2000CMH2OG
‡
100PSIG
.1
12FTH2O
.01
‡
200PSIG
.1
35FTH2O
†
.1
300PSIG
70FTH2O
†
1
.1
500PSIG
1
140FTH2O
.1
1000PSIG
1
230FTH2O
1
†
2000PSIG
1
480FTH2O
1
oz/in
700FTH2O
2
Res
1
50ZING
.1 1150FTH2O
1
‡
80ZING
.1
mmHg
Res
‡
240ZINA
150MMHGG
‡
1
‡
.1
240ZINVAC
260MMHGG
‡
1
‡
1
±240ZING
1
760MMHGA
1
‡
240ZING
†
1
760MMHGVAC
‡
1
480ZINA
1 ±760MMHGG
‡
‡
1
480ZING
760MMHGG
1
1
inHg
Res
1600MMHGA
1
6INHGG
‡
.01 1600MMHGG
1
10INHGG
Torr
‡
.01
Res
30INHGA
760TORRA
†
.1
1
30INHGVAC
.1
760TORRVAC
1
‡
‡
±30INHGG
‡
.1 1600TORRA
1
30INHGG
mbar
†
.1
Res
60INHGA
200MBARG
.1
‡
.1
60INHGG
.1
350MBARG
1
‡
120INHGG
.1
1000MBARA
1
200INHGA
1000MBARVAC
.1
‡
1
200INHGG
.1 ±1000MBARG
‡
1
400INHGG
1
1000MBARG
1 ±1000GCMG
600INHGG
1
2000MBAR
1
1000INHGG
2000MBARG
1
1
2000INHGG
bar
1
Res
atm
Res
1BARA
.001
1ATMA
.001
1BARVAC
‡
.001
1ATMVAC
±1BARG
‡
.001
‡
.001
±1ATMG
.001
1BARG
.001
‡
1ATMG
.001
2BARA
.001
2ATMA
.001
2BARG
.001
2ATMG
4BARG
.001
.01
4ATMG
.01
7BARA
.01
7ATMA
.01
7BARG
.01
7ATMG
.01
14BARG
.01
14ATMG
20BARG
.01
.01
20ATMG
.01
35BARG
.1
†
34ATMG
.1
70BARG
.1
†
70ATMG
.1
140BARG
.1
140ATMG
200BARG
.1
.1
200ATMG
.1
350BARG
1
†
340ATMG
1
†
Potentiometer
covers (3)
Standard:
plastic caps
TEST
Span
Test
NEMA 4X:
Cecomp Electronics / Made in USA
nylon screws
with o-rings
¼" NPT
3.38" Standard
3.5" NEMA 4X
C E CO MP
Div. of
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Standard accuracy: ±0.25% of full scale ±1 least significant
HA accuracy option: ±0.1% FS ±1 LSD, see ranges for
Sensor hysteresis:
Sensor repeatability: ±0.01% FS, included in accuracy
mmH
O
Res
2
2000MMH2OG
‡
1
Display
cmH
O
Res
2
3.5 digit LCD, 0.5" digit height (indicates to 1999)
200CMH2OG
.1
‡
3 readings per second nominal display update rate
350CMH2OG
‡
1
DRBL: Red LED backlight
1000CMH2OA
1
Controls
1000CMH2OVAC
1
‡
Non-interactive zero and span, ±10% range
1
‡
Output test adjustment:
1000CMH2OG
1
Retransmission zero and span: Internal potentiometers
2000CMH2OA
‡
1
1
Output Characteristics
kPa
Res
True analog output, 50 millisecond typical response time
20KPAG
‡
.01
-I version:
Current output, 4-20 mA DC, output drive
35KPAG
‡
.1
(compliance) determined by power source.
100KPAA
.1
See graph on other side.
100KPAVAC
.1
‡
-V version: Voltage output, 0-2 VDC into 5k ohm or greater.
±100KPAG
‡
.1
Output is 0-1-2 V with ±15 psi sensor.
100KPAG
.1
200KPAA
.1
Test Function
200KPAG
.1
Front panel TEST button, when depressed sets loop current
400KPAG
1
and display to output test level, independent of pressure
700KPAA
1
input, to allow testing of system operation.
700KPAG
1
Power
1400KPAG
1
8-24 VAC 50/60 Hz or 9-32 VDC
2000KPAG
1
Use with WMPSK 115 VAC/12 VDC power supply
MPa
Res
Gauge is on whenever power is applied
1.4MPAG
.001
Designed for continuous operation
2MPAG
.001
DR:
30 mA maximum
3.5MPAG
†
.01
DRBL:
Approximately 40 mA maximum
7MPAG
.01
14MPAG
.01
Weight
20MPAG
.01
9.5 ounces (approx.)
35MPAG
†
.1
Shipping wt. 1 pound (approx.)
g/cm
2
Res
Housing
200GCMG
.1
‡
DPG1000DR: Extruded aluminum case, epoxy powder coated,
350GCMG
‡
1
ABS/ polycarbonate bezel (aluminum bezel optional), front and
1000GCMA
1
rear gaskets, polycarbonate label
1000GCMVAC
‡
1
F4DR: ABS/polycarbonate NEMA 4X case, rear gasket, poly-
1
‡
carbonate label
1000GCMG
1
2000GCMA
Connection, Material, Media Compatibility
1
2000GCMG
1/4" NPT male fitting, 316L stainless steel
1
kg/cm
Res
All wetted parts are 316L stainless steel
2
Compatible with most liquids and gases
1KGCMA
.001
1KGCMVAC
‡
.001
Overpressure
±1KGCMG
.001
‡
2 X pressure range for 3 psi to 2000 psi sensors
1KGCMG
.001
5000 psig for ranges using 3000 psig sensor
2KGCMA
.001
7500 psig for ranges using 5000 psig sensor
2KGCMG
.001
Vacuum service: 15 psia, ±15 psig, 15 psig, 30 psia,
4KGCMG
.01
7KGCMA
.01
Burst Pressure
7KGCMG
.01
4 X sensor pressure rating or 10,000 psi, whichever is less
14KGCMG
.01
20KGCMG
.01
Environmental
35KGCMG
.1
Storage temperature:
†
70KGCMG
.1
Operating temperature:
140KGCMG
.1
Compensated temperature:
200KGCMG
.1
350KGCMG
1
†
2.88"
Standard
Zero
Allow 1"
clearance
0.75"
1.65"
A
P
BSOLUTE
ROCESS
digit
availability
±0.015% FS, included in accuracy
0-100% range
100 psig, 100 psia, 200 psig sensors
–40 to 203°F (–40 to 95°C)
–4 to 185°F (–20 to 85°C)
32 to 158°F (0 to 70°C)
3 ft long 4-conductor
3.0"
22 AWG shielded
cable with stripped
and tinned wire ends
NEMA 4X
Allow 1"
clearance
Turn at hex
fitting only
2.0"
1220 American Way Libertyville, IL 60048
I
NSTRUMENTS
DPG1000DR, F4DR
O ±0.25% Test Gauge Accuracy
O 316 Stainless Steel Wetted Parts
O Low Voltage Powered
O 4-20 mA or 0-2 V Analog Output
O Output Test Function
F4DR
Quick Link
cecomp.com/trans
How to Specify
DPG1000DR range - output - options
DPG1000DRBL range - output - options
F4DR range - output - options
F4DRBL range - output - options
Range—see table at left
psi = PSI
torr = TORR
inHg = INHG
mmH
O = MMH2O
2
oz/in
= ZIN
kg/cm
= KGCM
2
2
inH
O = INH2O
g/cm
= GCM
2
2
ftH
O = FTH2O
kPa = KPA
2
mmHg = MMHG
MPa = MPA
G = gauge reference pressure
VAC = gauge reference vacuum
A = absolute reference
Range codes listed as 2, 20, 200, or 2000 display 1.999, 19.99,
199.9, or 1999 respectively.
Output
Specify:
I
4-20 mA
V
0-2 V
Options—add to end of model number
High accuracy, ±0.1% FS ±1 LSD. See table at left for
HA
availability.
PM
Panel mount, 4.1" x 4.1", n/a NEMA 4X
CC
Moisture resistant circuit board conformal coating
CD
Calibration data; 5 test points and date
NC
NIST traceability documentation, 5 points and date
Accessories—order separately
WMPSK Wall mount power supply kit, 115 VAC/12 VDC
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. Use for food vacuum packaging applica-
tions. 303 SS body, 100 micron 304 SS screen.
cecomp.com
800-942-0315
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
DPG1000DR
NEMA 4X
Housing
Type
Standard housing
Standard housing,
backlit display
NEMA 4X housing
NEMA 4X housing,
backlit display
mbar = MBAR
bar = BAR
cmH
O = CMH2O
2
atm = ATM
© 08-18