Auber Instruments AT210 Manual de instruções - Página 2

Procurar online ou descarregar pdf Manual de instruções para Termómetro Auber Instruments AT210. Auber Instruments AT210 4 páginas. Digital
Também para Auber Instruments AT210: Notas sobre a utilização (2 páginas), Manual de instruções (3 páginas), Manual de instruções (3 páginas)

AUBER INSTRUMENTS
2. Check the peak value. Press Down key once to show peak
temperature, and the MAX LED will light up; press Down key again to
show the time stamp, i.e., the time when peak temperature is reached
(count from the start of the timer, unit is seconds), and the MAX LED
will flash. If the temperature continues rising, a new peak temperature
and its time will be written to the memory to replace the previous peak
value. Please see Section 9 for details.
3. Dimmer the display brightness. Press Up key to toggle the display
brightness between the dimmed display and the normal bright display.
The brightness of the dimmed display is determined by the parameter
"brit" described in the later section of the manual (see Section 7).
6. Alarm setting (code 0001)
The meter has two programmable alarms that can be set to turn on the
buzzer at specified temperatures. The alarm1 is controlled by
parameters AH1 and AL1; the alarm2 is controlled by AH2 and AL2.
These parameters can be accessed by code 0001 (see Table 1).
Table 1. Alarm parameters.
Symbol
Name
Description
AH1
AH1
Alarm1 on temperature
AL1
AL1
Alarm1 off temperature
AH2
Alarm2 on temperature
AH2
AL2
AL2
Alarm2 off temperature
END
END
Exit
AH1 and AH2 are called alarm on temperature, i.e. the temperatures at
which the Alarm1 and Alarm2 will turn on; AL1 and AL2 are called alarm
off temperature, i.e. the temperatures at which the Alarm1 and Alarm2
will turn off. When AH1 > AL1(or AH2 > AL2), the alarms are set as a
high limit alarm. When AH1 < AL1(or AH2 < AL2), the alarm is set as a
low limit alarm. For a high limit alarm, the alarm buzzer should be on
once the temperature rise above AH, and it should be off only when the
temperature drops below AL. In contrast, for a low limit alarm, the alarm
buzzer should be on when the temperature drops below AH, and it
should be off only when the temperature rises above AL.
For example, when the temperature unit is set to F (Fahrenheit), AH1 =
900, and AL1 = 800, the buzzer will go off once the measured
temperature is higher than 900° F; the buzzer will stop when the
temperature drops below 800° F. In another case, when AH2 = 180, AL2
= 185, the alarm2 is set as a low limit alarm. If the temperature drops
below 180° F, the buzzer will go off (on); when the temperature rises
above 185° F, the buzzer will stop.
User can press the Shift key (">") to mute the buzzer. The alarm will buzz
again if the alarm temperature moves out of the alarm zone (between
AH and AL) and re-enters it again. For instance, AH1 = 900, AL1 = 800,
current temperature is higher than 900, and the alarm is muted. In this
case, the temperature has to drop below 800 and rise above 900 again
to trigger the alarm buzzer. To disable the alarms, set AH and AL to the
same value, i.e., AH1 = AL1 and AH2 = AL2.
Steps to set an alarm are described below as well as in the flow chart in
Figure 4:(1) Press SET key, change the code to 0001, and press SET
again to enter alarm setting mode. (2) Press Up and Down key to select
a parameter (AH1, AL1, AH2, or AL2). (3) Press SET to view the value
2020.09
Sett ing Range
Initial
Note
-1999 ~ 9999
480
-1999 ~ 9999
479
-1999 ~ 9999
250
-1999 ~ 9999
250
of the parameter.(4) Use Shift, Up, and Down key to change the value.(5)
Press SET key to confirm the value and exit from this parameter.(6)
Press Up or Down key to select the new parameter.(7) To exit this mode,
press SET key when "End" is displayed. The procedure is showed in the
flow chart in Figure 4.
Oper ation Mode
Enter Code
SET
SET
0001
XXXX
SET
Figure 4. Steps of setting alarms. In this example, alarm1 is set as a high
limit alarm at 480; alarm2 is disabled.
7. Alarm setting (code 0001)
The parameters for system configuration are listed in Table 1.
Table 1. System configuration parameters.
Note 1. Inty (Input sensor type). This parameter defines the sensor type
that is used for the thermometer. If you purchased the thermometer
bundle, it is already set for the probe included in the package. You don't
need to change it.
Note 2. PSb (input offset). This is to set an offset value to compensate
the error produced by the sensor. For example, if the meter displays 5
ºC when probe is in ice/water mixture, setting PSb= -5 will make the
meter display 0 ºC. You can use shift key (second key from the left) to
shift to the first left digit, then press down key once to enter minus sign.
Note 3. FILt (digital Filter). If the temperature reading fluctuates due to
input noise, a digital filter can be used to smooth the input signal. "FILt"
may be configured in the range of 0-3. Higher the value, stronger the
filtering effect. A strong filter increases the stability of the readout, but
causes longer delay in responding to the temperature change. Set
FILt=0 will disable the filter.
Note 4. C-F (temperature unit). Set to C if you want to display
temperature in Celsius. Set to F for Fahrenheit.
Note 5. Brit (display dimmer). This parameter sets the brightness level
of the dimmed LED display. You can switch the LED brightness level
between the dimmed and the normal using UP key. However, the normal
brightness level of the LED display is equal to level 4. So if Brit is set to
4, you won't see and change in the brightness of LED display when
pressing UP key.
Note 6. End. Exit the parameter setting mode.
WWW.AUBERINS.COM
Parameter Display
Buzzer On Temp
SET
0480
AH1
Buzzer Off Temp
SET
0479
AL1
Buzzer On Temp
SET
0250
AH2
Buzzer Off Temp
SET
0250
AL2
END
SET
SET
SET
SET
P2/4