should be loaded on the platform. If you have to change this weight, you can press
or
the flashing position and press
plus 1, and press
g) After making sure the weight on the platform is"xx.xkg", press
confirm so that the scale begins the "xx.xkg"calibration. In the process of"xx.xkg"
calibration, "xx.xkg" flickers rapidly on the display window;keep the platform calm,
stable, and without other interference factors which will influence the weighing
performance. If the weighing remains stable for more than 2s, the scale will finish
"xx.xkg"calibration and save the calibrated parameters, The indicator will return to
the weighing mode

4.5.2 Standard Two Point Weight Calibration

a) At weighing mode, press and hold
b) Press
c) Press
d) After the scale begins to calibrate, the window will display"Load"and"0.0kg"
alternately;
e) After you confirm that there is nothing on the platform, press
confirm and begin the zero calibration when the window displays"0.0kg". In the
process of zero calibration,"0.0kg"flickers rapidly on the display window;please keep
the platform calm, stable, and without other interference factors which will influence
the weighing performance. If the weighing keeps stable exceeding 2s, the scale will
finish zero calibration automatically, and then it will begin the next calibration point;
f) Repeat the step d) and step e) above,calibrate the half capacity range point and the
Full capacity range point orderly, save the calibrated parameters automatically and
return to the weighing mode.
4 |
A n y l o a d D D - K C 1 M a n u a l V ( 1 6 0 3 )
button to enter the modify menu; press
button to finish the modification
button to cycle the display until the window displays"Line";
or
button to begin calibration procedure;
button to make the flashing position's number
button until the window displays"------";
button to change
button to