If levels do not adjust appropriately
according to patient's walking pace, tap
Controls>Settings>Step Rate and adjust the
Medium and High Step rates as needed. Tap
Save and repeat confirmation again by asking the
patient to walk slow, medium and fast paces to
make sure vacuum levels adjust correctly.
Outcome Report
The iQ Vacuum device is designed to collect
and provide both performance and qualitative
outcomes of the prosthesis user.
•
PEQ Questionnaire for patient satisfaction and
utility
•
Cadence Variation and Step Count