CABLE CONNECTORS

Apogee sensors offer cable connectors to simplify the
process of removing sensors from weather stations
for calibration (the entire cable does not have to be
removed from the station and shipped with the
sensor).
The ruggedized M8 connectors are rated IP68, made
of corrosion-resistant marine-grade stainless-steel,
and designed for extended use in harsh
environmental conditions.
Instructions
Pins and Wiring Colors: All Apogee connectors have
six pins, but not all pins are used for every sensor.
There may also be unused wire colors inside the
cable. To simplify connection to a measurement
device, the unused pigtail lead wire colors are
removed.
If a replacement cable is required, please contact
Apogee directly to ensure ordering the proper pigtail
configuration.
Alignment: When reconnecting a sensor, arrows on
the connector jacket and an aligning notch ensure
proper orientation.
Disconnection for extended periods: When
disconnecting the sensor for an extended period of
time from an installation, protect the remaining half
of the connector still on the station from water and
dirt with electrical tape or other method.
Tightening: Connectors are designed to be firmly
finger-tightened only. There is an O-ring inside the
connector that can be overly compressed if a wrench
is used. Pay attention to thread alignment to avoid
cross-threading. When fully tightened, one to two
threads may still be visible.
WARNING: Do not tighten the connector by twisting
the black cable, only twist the metal connector.
Inline cable connectors are installed 30 cm from the
sensor head (pyranometer pictured).
A reference notch inside the connector ensures
proper alignment before tightening.
When sending sensors back for recalibration, only send
the section of cable that is hard-wired to the sensor. The
section of cable with the pigtail is not required.
Finger-tighten firmly.