Earth-bond resistance Measurement
Function
Range
40.00Ω
Earth-bond
400.0Ω
Resistance
4000Ω
40.00kΩ
*<1.00Ω add 3dgt
Open Circuit Test Voltage : >4.0V,<8V
Short Circuit Current : >200.0mA
Live Circuit Detection: if≧2V ac/dc at inputs, test inhibited.
Insulation resistance Measurement
Function
Range
4.000MΩ
40.00MΩ
400.0MΩ
Insulation
4000MΩ
Resistance
4.1GΩ~20.0GΩ
* Above specifications only apply when high quality silicone
leads with test clips are being used with no hands touch.
Test Voltage vs. Maximum resistance range :
50V/50.0MΩ,100V/100.0MΩ,250V/250.0MΩ,500V/500MΩ and
1000V/20.0GΩ.
Test Voltage vs. Minimum resistance (with test
current=1mA) :
50V/50kΩ,100V/100kΩ,250V/250kΩ,500V/500kΩ and
1000V/1MΩ.
Test Voltage Accuracy : -0%,+20%
Short Circuit Test Current:1mA(nominal)
Auto discharge function : discharge time<1 sec for C≦1
Maximum Capacitive load : Operable with up to 1
Live Circuit Detection : if≧30V ac/dc at inputs, test inhibited
Accuracy
±(1.5%+5dgt)*
±(1.5%+3dgt)
Accuracy *
±(1.5%+5dgt)
±(3%+5dgt)
±(10%+3dgt)
µ
F load
µ
15
F