Shodex RI-101
6 Troubleshooting
When the solution to a problem cannot be found easily, a recording of the
communication between Clarity and the detector will significantly help
DataApex support.
The data recording can be enabled by adding or amending the
COMMDRV.INI file in the Clarity installation directory (C:\CLARITY\CFG
by default). The file can be edited in any text editor (e.g. Notepad). The
following section should be edited or added:
[COM1]
echo=on
textmode=on
filename=CommDrvCOM1_%D.txt
reset=off
Note:
Instead of COM1, type the communication port used to communicate with
the Shodex RI-101 detector. This port number is displayed when the Det
Status button in the
Note:
%D (or %d) in the filename parameter means that the log will be created
separately for each day. The reset=off parameter disables deleting the
content of the log each time the station is started during the same day.
The created *.TXT files will be of great help in the diagnosis of not
documented errors and communication issues.
Method Setup - Acquisition
- 14 -
6 Troubleshooting
dialog is invoked.