1. Unpack the machine and preserve the original package which shall be used for any
returns or repairs., place the appliance on a stable surface capable of supporting at
least 70 Kg.
2. VENTILATION: In the case the sterilizer should be placed close to walls or be fitted-in,
ensure observance of the minimum ventilation dimensions, i.e. at least 5cm of space all
round to allow air circulation around the machine.
3. Check that the appliance is positioned slightly inclined towards the back. To this end,
pour a small quantity of water (10cc) in the chamber and ensure it flows towards the
bottom.
Should the inclination need to be changed, regulate the adjustable feet.
Connect the power cable located at the back of the appliance, open the cover below the
front display and press the green power switch
The appliance must be connected to a grounded mains socket in compliance with IEC
Standards, with a mains voltage of 230 Volts and minimum capacity of 16 Amps.
WARNING!!
♦ Do not connect other appliances to the same socket.
♦ Should it be necessary to use an adapter, ensure its capacity is higher than the
maximum sterilizer power ( 2,090 watts).
4. Switch on the main green switch located under the cover
below the touch-screen display
5. The display shows a short self test
Rev. 00 – 06/08/2014