AUSTIN SCIENTIFIC M600-HV Manual - Página 11
Procurar online ou descarregar pdf Manual para Compressor de ar AUSTIN SCIENTIFIC M600-HV. AUSTIN SCIENTIFIC M600-HV 13 páginas.
6
MAINTANENCE AND SERVICE
Before any service verify that the power has been removed from the
system
6.1
ABSORBER REMOVAL
The service interval of the absorber of the M600 is 20,000 hours (27.4
months).
The M600 absorber has been placed at the front of the unit, behind a
removable cover, to facilitate replacement. Absorber replacement
should take about 15 minutes. This is ideal for rack and rail mounted
compressor units.
6.2
HELIUM PRESSURE ADJUSTMENTS
The helium charge can be increased or decreased by using the ¼" self-
sealing connector on the front of the unit, this connection is marked
by a white washer.
Never charge the compressor above a static pressure of 250psi
(17.2bar)
7.
TROUBLE SHOOTING
For any trouble or questions installing and operating the M600 unit
please contact the service department of Austin Scientific at:
Austin Scientific
Phone
Fax
(512) 441-6893 x 112
(512) 443-6665