ACCU-CAL™ 160 Radiometer User Guide

Setup Menu

To enter the setup menu, switch the device on and wait for the welcome screen (Figure 2) to
disappear. The graph mode or text mode screen will appear. Press the button in the middle to
enter the setup (Figure 3).
Figure 2. Welcome Screen
Figure 3. Text Mode Screen - Push the Middle Button to Enter Setup
Once on the setup screen (Figure 4), you can toggle the units for irradiance and dose (per m² or
per cm²), the temperature scale (°C or °F), the sample rate (low (25 Hz), or high (2,500 Hz)) the
lamp type of your application for spectral correction (press the right button several times to
select your lamp type. If your specific lamp model is not listed, select "unknown").