DayTronic 3770 Manual de instruções - Página 6
Procurar online ou descarregar pdf Manual de instruções para Instrumentos de medição DayTronic 3770. DayTronic 3770 16 páginas. Dc strain gage panel meter
Model 3770
2
C
O N N E C T I O N S
J1 – Shunt Command Connections. Connection is used to activate the positive or negative
Shunt calibration command from the rear panel. When the + or – R CAL terminal is
connected to LOGIC COM via a switch or relay, the installed Shunt resistor will be
connected across + Signal and + Sense or – Sense depending on the Shunt polarity chosen
and the CAL SENSE connection. CAL SENSE line is normally connected to + Signal for
proper signal activation.
Fig. 4
J2 - HOLD & TARE Connections.
or "TARE" features of the unit are enabled and are controlled by an external switch, PLC or
relay. The LOGIC COM signal is used to activate the analog HOLD or Analog TARE
(offset) to the meter's display and analog output signals present on J3.
Fig. 5
J1 Logic Controls
HOLD IN
TARE IN
TARE ENABLE
LOGIC COMMON
N/O
Switch
SHUNT Command
Connections
Connection is used when external control of the "HOLD"
Push Button, Contact Relay,
Switch.....
Activates Analog HOLD when connected to LOGIC COMMON
Activates Analog TARE when connected to LOGIC COMMON
Enables remote TARE and disables front panel TARE when connected to LOGIC COMMON
Connection to enable rear panel features
DC STRAIN GAGE PANEL METER
J1 - SHUNT Activation
Connections
Connecting LOG COMMON to the R CAL + or
R CAL - terminal will activate the installed
shunt resistor (internal or external) across the
connected four arm bridge sensor for
simulated calibration & verification check of
the sensor's electrical characteristics.
Page 5