DayTronic 5D78 Manual de instruções - Página 5

Procurar online ou descarregar pdf Manual de instruções para Instrumentos de medição DayTronic 5D78. DayTronic 5D78 20 páginas. Panel meter, ac strain gage conditioner module

DayTronic 5D78 Manual de instruções
1.
G
A
ENERAL
T
3000PLUS P
HE
ANEL
Incorporating Daytronic's 5D Series Signal Condition-
er Modules, the 3000PLUS Panel Meter is a field-scal-
able indicator featuring operator-programmable signal
processing and PC/PLC communications. With a
durable front panel and secure screw terminals for all
power and I/O connections, this mechanically and elec-
trically rugged instrument is ideal for pump, motor,
hydraulic, and other high-noise monitoring applications.
Accepting the fully conditioned output of any standard
plug-in 5D module*, the 3000PLUS maintains signal
integrity to deliver accurately scaled analog output,
while sampling all data at 16-bit resolution. The data dis-
play provides selectable digital filtering for even greater
readout stability.
The operator can easily select any of three separate
output channels for display, as explained in Section 1.D:
• the meter's standard ±5-VDC scaled output
(Channel 1), representing measured engineering
units (after calibration)
• the "auxiliary" DAC output (Channel 2), which
serves to monitor and process Channel 1, and to
generate Channel 3
• the "scaled voltage" output (Channel 3), which is
continuously proportional to the reading of the auxil-
iary output (Channel 2). Channel 3 may be set to a
full scale of either ±5 or ±10 VDC, and is available
from the rear of the 3000PLUS as both voltage and
4-20 mA output (see Section 2.D).**
The auxiliary analog output (Channel 2) is used for
• HI/OK/LO limit monitoring with selectable hystere-
sis windows, front-panel annunciation, and relay out-
puts for local process control (see Section 5.E for
full details)
• high-speed positive or negative peak capture with
TTL-level "have peak" output, selectable "backout"
and "peak defeat" thresholds, and user-adjustable
leak rate (Section 5.B)
• a signal hold to allow captured peaks and other
values to be transferred to computer for processing
(Section 5.C)
• automatic application of a desired tare offset (Sec-
tion 5.D)
* "V" and "S" models may not be used with the 3000PLUS.
** The ±5-VDC output of the installed 5D78 conditioner (prior to
A/D conversion) is also available from the rear of the meter.
3000PLUS
D
ESCRIPTION AND
M
ETER
5D78 I
WITH
NSTRUCTION
1. I
NTRODUCTION
S
PECIFICATIONS
Separate logic inputs provide external control of peak
capture, signal hold, tare application, and the release of
latched limit violations.
As explained in Sections 3 and 4 of this manual, you
can quickly set up the 3000PLUS either via the simple
front-panel button menu or via the configuration soft-
ware supplied with the unit. Operator-entered ranges,
filters, calibration points, and other setup parameters
are always specific to the installed 5D Series condition-
er.
A standard RS232 interface operating at a fixed rate of
19.2K baud allows connection of an external PC for
instrument configuration and on-line monitoring. It may
also be used for direct communication of simple ASCII
mnemonic commands both for run-time reconfiguration
and for interrogation of current data and setup values.
Employing the run-time version of Microsoft
2000, the 3KP CONFIGURATOR software supplied
with the Model 3000PLUS makes short work of meter
setup. Communicating via the RS232 link, the Configu-
rator lets you define, store, edit, download, upload, and
manage any number of "configurations" for your
3000PLUS. As explained in Section 4, "off-line" configu-
rations can be easily created and downloaded, or you
can use the "Live Output Window" to view and modify
the present configuration of the connected 3000PLUS
instrument on a purely run-time basis.
The Configurator also lets you perform selected run-
time operations, including
• both "absolute" and two-point (deadweight) calibra-
tion
• viewing any of the meter's three "live" analog out-
puts and adjusting it as desired
• applying a signal hold command
• releasing any and all latched limits
• sending standard mnemonic commands to the
meter
• activating the shunt resistor for verification of sensor
calibration data.
T
M
5D78 AC S
HE
ODEL
G
C
M
AGE
ONDITIONER
Delivering a high-level filtered analog output of ±5 VDC,
the Model 5D78 is a single-channel strain gage condi-
tioner of phase-sensitive carrier-amplifier design.
While intended primarily for applications involving trans-
former-coupling to the transducer bridge (as with con-
ventional rotary-transformer torque sensors), the
Model 5D78 can also be used when high sensitivity is
required or where the electrical environment is espe-
M
ANUAL V
®
Access
TRAIN
ODULE
3KP78 - 1.1
. SB.3