Automation Devices 6472.3 Manual - Página 9

Procurar online ou descarregar pdf Manual para Eletricidade industrial Automation Devices 6472.3. Automation Devices 6472.3 10 páginas.

Model 6476.3
Constant Feed Rate Sensor
INSTALLING THE CFR SENSOR
Note: Failure to adequately prepare the
feeder's surface properly may result in a
Constant Feed Rate (CFR) sensor that will not
bond to the feeder. The sensor should not be
mounted until step C-6.
1.375"
P/N 123-215
Fig. 2 The arrow shows the direction of vibration
which is at a right angle to the spring pack.
A. ORIENT THE SENSOR so that its sensitive
axis is in the same direction as the vibration of the
feeder. The double-ended arrow in figure 1 shows
the sensor's sensitive axis. Align the sensitive axis of
the sensor in the same direction as the vibration (see
figure 2). The sensor must be oriented correctly for
proper operation.
B. CHOOSE A LOCATION for mounting the sensor
on the feeder that is smooth and that will allow the
adhesive on the sensor to bond. Avoid mounting the
sensor over ridges and bumps which can reduce the
ability of the adhesive to stick to the feeder. The
correct location will also have enough space for the
sensor's cable to hang straight down without touching
anything else.
C.
SURFACE PREPARATION of the feeder is
crucial for proper bonding between the sensor and
the feeder. Please follow these steps completely.
1) The feeder should be kept between 70 and 100
F for ideal tape application.
2) Clean a three and one-half inch circular area with
a solvent like isopropyl alcohol that will not leave a
Sensitive
residue.
As a rule of thumb, the area can be
Axis of
considered clean when after cleaning the area with
Vibration
a solvent-saturated, white paper-towel, the towel is
as clean as it was before wiping.
3) Using a good amount of pressure, polish the
cleaned, circular area of the feeder using a scratch
Fig. 1
pad or steel wool. Repeat step 2, and then go to
Actual Size
step 4.
4) Wipe the cleaned surface with an alcohol wipe or
with a 50/50 isopropyl alcohol/water combination.
5) Dry the surface thoroughly using a low lint cloth or
a clean paper towel.
6) Remove the vibration sensor from its protective
packaging. Remove the liner from the adhesive
backing. Avoid touching the tape. Align the sensor
as shown in figures 1 and 2. Apply the vibration
sensor to the prepared area of the feeder. Press
the sensor very firmly onto the feeder surface for at
least 10 seconds.
7) Allow the vibration sensor at least 20 minutes to
cure before operation. Note: It takes 72 hours for
the adhesive to fully cure at 70 F.
Alternatively, #8 or M4 screws can be used to mount
the sensor to the feeder. The mounting holes are
1.375" (3.49cm) apart.
D. ROUTE THE SENSOR CABLE to protect it from
strain due to vibration. The cable that attaches to the
sensor will not break from normal vibration; however,
some care should be used when routing the sensor
cable from the sensor to the control.
should hang straight down from the sensor without
touching the feeder bowl or anything else. Then,
the sensor cable should curve towards the power
control with a bend radius larger than 3 inches.
Use a cable tie and an adhesive-backed mount to
attach the sensor cable to the side of the drive
base. See Figure 2. Clean the mounting area
before applying the adhesive-backed mount.
E. CONNECT THE SENSOR to the control. If
needed, connect the sensor cable's brown wire to
"+ACCEL" on TB2. The blue wire connects to the
"–ACCEL" on TB2.
F. SELECT THE SENSOR in the software
settings of the control menu. See the Control
Menu Layout page for a visual layout of the
program menu.
1) Press and hold the "Enter" key to enter the
main menu.
2) With "Power Settings" displayed, press the
"Enter" key to get into the submenu.
3) Arrow "Down" to select the "Amplitude Source"
submenu.
4) Press the "Enter" key, and Arrow "UP" to select
"Auto Tracking" from the adjustments.
5) Press the "Back" key twice to get back to the
main menu.
6) Next, arrow "Down" to the "Frequency
Settings" selection.
7) Press the "Enter" key to get into the submenu.
8) Arrow "Down" to select the "Frequency Mode"
submenu.
9) Press the "Enter" key, and Arrow "UP" to select
"Auto Tracking" from the adjustments.
10) Press the "Back" key three times to get back
to the normal running display.
G. PERFORM AUTO SCAN of the vibratory
feeder. See the Control Menu Layout page for a
visual layout of the program menu.
1) Press and hold the "Enter" key to enter the
main menu.
2) Arrow "Down" to select the "Frequency"
submenu.
3) Press the "Enter" key, and Arrow "UP" three
times to select "Auto Scan" from the adjustments.
4) Press the "Enter" key to perform an automatic
frequency scan.
Warning: Avoid dropping the sensor on a
hard surface. Damage could occur.
Automation Devices, Inc.
The cable
E-mail: [email protected], autodev.com
814-474-5561