Ordering Information
This is only for reference.
For selecting the specified model, follow the Autonics website.
KRN100
-
❶ ❷ ❸ ❹
❶ Input channel
02: 2 CH (universal input card × 1)
04: 4 CH (universal input card × 2)
06: 6 CH (universal input card × 3)
08: 8 CH (universal input card × 4)
10: 10 CH (universal input card × 5)
12: 12 CH (universal input card × 6)
❷ Digital input
0: None
1: 6 (digital input card × 1)
2: 12 (digital input card × 2)
❸ Alarm transistor output
0: None
1: 6 (transistor alarm output card × 1)
2: 12 (transistor alarm output card × 2)
Manual
For proper use of the product, refer to the manuals and be sure to follow the safety
considerations in the manuals.
Download the manuals from the Autonics website.
Software
Download the installation file and the manuals from the Autonics website.
■ DAQMaster
It is the comprehensive device management program for Autonics' products, providing
parameter setting, monitoring and data management.
Product Components
• Product
• Instruction manual
• Recording paper
• USB memory
Sold Separately
• Universal input card: KRN-UI2
• Transistor alarm output card: KRN-AT6
• Transmitter power output card: KRN-24V3
Specifications
Series
KRN100
LCD type
STN Graphic LCD
Resolution
320 × 120 pixel
Brightness adjustment
4-level (OFF / Min / Standard / Max)
Backlight
White LED, 2-level (Temp / Always)
No of input channel
2 / 4 / 6 / 8 / 10 / 12 CH model (2 CH / universal input card)
Universal input
Please refer to 'Input / Output' for detailed information about universal input.
1 to 4 CH: 25 ms / 125 ms / 250 ms, 5 to 12 CH: 125 ms / 250 ms
Sampling cycle
01)
(thermocouple (TC) - R, U, S, T: ≥ 50 ms)
Graph mode recording
10, 20, 40, 60, 120, 240 mm / H
speed
Recording speed
F.S. ± 0.5 %
accuracy
1 to 3600 sec (inner log file is saved at 1 sec interval)
Saving cycle
Internal memory
512 MB
External memory
USB memory max. 32 GB
02)
113 mm × 9 m
Recording paper
Normal printing is available after going and returning printing maximum 5
Ink cartridge
times within 7 days after opening the unit
Ink dry time
≤ 15 minutes
01) Internal sampling cycle is average movement filter and alarm output operation unit time.
02) USB memory is included in the box. If you use USB memory you purchased separately, it could not be recognized.
Power supply
100-240 VACᜠ 50 / 60 Hz
Allowable voltage range 85 to 110 % of rated power supply
Power consumption
≤ 23 VA
Between power terminals and case: 2500 VACᜠ 50 / 60 Hz for 1 minute
Dielectric strength
(except Ethernet and USB device)
Vibration (conveying
10 to 60 Hz 4.9 m / s
and storing)
Vibration (operating)
10 to 60Hz 1 m / s
Insulation resistance
≥ 20 MΩ (500 VDCᜡ megger)
Noise immunity
± 2 kV square wave noise (pulse width 1 ㎲) by noise simulator
Time accuracy
Within ± 2 min / year (available up to 2100 year)
Protection structure
IP50 (front part, IEC standard)
0 to 50 ℃, storage: -20 to 60 ℃
Ambient temperature
(without the ink cartridge, no freezing or condensation)
Ambient humidity
35 to 85 %RH, storage: 35 to 85 %RH (no freezing or condensation)
Approval
ᜢ ᜣ ᜫ
Unit weight (packaged) ≈ 1.7 to 2.0 kg (≈ 2.4 to 2.7 kg)
-
❺ ❻
-
❹ Alarm relay output
0: None
1: 4 (relay alarm output card × 1)
2: 8 (relay alarm output card × 2)
3: 12 (relay alarm output card × 3)
❺ Transmitter power output
0: None
1: 3 (transmitter power output card × 1)
2: 6 (transmitter power output card × 2)
3: 9 (transmitter power output card × 3)
4: 12 (transmitter power output card × 4)
❻ Communication output
0: None
1: RS485 / Ethernet / USB
(communication output card × 1)
• Ink cartridge
• Basic connector × 2
• Bracket × 2
(the no. of additional
connectors depends on
the input/output card.)
• Digital input card: KRN-DI6
• Relay alarm output card: KRN-AR4
• Communication output card: KRN-COM
X, Y, Z in each X, Y, Z direction for 1 hour
2
X, Y, Z in each X, Y, Z direction for 10 minutes
2
Input / Output
Connect or disconnect the input / output card when the product is turned off.
■ Universal input card (KRN-UI2)
• Input specifications
0
S
RTD
Thermocouple
Voltage
Analog
Current
If sensor input line is longer, it is recommended to use shield cable to reduce noise.
• Input impedance
RTD, thermocouple,
voltage (mV)
Voltage (V)
Current
• Display accuracy
Input type
RTD
Thermocouple
Analog
• Resolution: 16 bit
■ Digital input card (KRN-DI6)
Non-contact input
Contact input
■ Alarm transistor output card (KRN-AT6)
NPN Open Collector, 12-24 VDC / ≤ 30 mA
■ Alarm relay output card (KRN-AR4)
Capacity
Mechanical life cycle
Electrical life cycle
■ Transmitter power output card (KRN-24V3)
24± 2 VDCᜡ, total 3 CH, per 1 CH ≤ 30 mA (built-in over current protection circuit)
■ Communication output card (KRN-COM)
RS485
EEPROM life cycle
Ethernet
USB Device
RS422 / 485 and Ethernet communication outputs cannot be used at the same time.
The front USB Device port is only for data backup and rear USB device port is only for
parameter setting.
Input / Output Card
Universal input card
[KRN-UI2]
-
+
-
+
• Each channel is isolated and
dielectric strength is 500V.
Alarm relay output card
[KRN-AR4]
Inside
JPt100Ω, DPt100Ω, DPt50Ω, Cu100Ω, Cu50Ω (supplied current 420 ㎂)
B, C (W5), E, G, J, K, L, L (Russia), N, P, R, S, T, U
± 60 mV, ± 200 mV, ± 2 V, 1-5 V, ± 5 V, -1 V-10 V
0.00-20.00 mA, 4.00-20.00 mA
≥ 2 MΩ
≥ 150 kΩ
51 Ω
Temperature
Display accuracy
± 0.1 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
Room temperature
• Cu50Ω (-200 ≤ T ≤ 200): ± 1.0 ℃
range (25 ℃ ± 5 ℃)
• DPt50Ω (-200 ≤ T ≤ 500): ± 1.5 ℃
± 0.2 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
Out of room
• 500 to 850 ℃ of all RTDs: ± 0.5 % of PV value ± 1 digit
temperature range
• Cu50Ω (-200 ≤ T ≤ 200): ± 2.0 ℃
• DPt50Ω (-200 ≤ T ≤ 500): ± 3.0 ℃
± 0.1 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
• R, S, C, G (0 ≤ T ≤ 100): ± 4.0 ℃
Room temperature
• U, T (-200 ≤ T ≤ -100): ± 3.0 ℃
range (25 ℃ ± 5 ℃)
• U, T (-100 ≤ T ≤ 400): ± 2.0 ℃
• Below 400 ℃ of B: there is no accuracy standards.
• Below -100 ℃ of all thermocouples: ± 0.3 % F.S. ± 1 digit
Out of room
± 0.2 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
temperature range
Room temperature
± 0.1 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
range (25 ℃ ± 5 ℃)
Out of room
± 0.2 % F.S. ± 1 digit (warm-up time: ≥ 30 minutes)
temperature range
ON: residual voltage ≤ 1 VDCᜡ, OFF: leakage current ≤ 0.1 mA
ON: ≤ 1 kΩ , OFF: ≥ 100 kΩ, short-circuit: ≈ 4 mA
250 VACᜠ 3 A, 30 VDCᜡ 3 A, 1 Form A (resistive load)
≥ 50,000,000 operations
≥ 100,000 operations (250 VACᜠ 3 A, 30 VDCᜡ 3 A)
Modbus RTU (It is recommended to use shielded cable over AWG 24.)
≈ 1,000,000 operations (Erase / Write)
IEEE802.3(U), 10 / 100 BASE-T (Modbus TCP)
USB V2.0 Full Speed (Device Control)
Digital input card
[KRN-DI6]
Transmitter power output card
[KRN-24V3]
• Each 24 V output is isolated
individually.
Alarm transistor output card
[KRN-AT6]
Inside
• AL1,2,3 and AL4,5,6 are isolated.
Communication output card
[KRN-COM]