CLEANING INSTRUCTIONS
CUSHION
1. Use germicidal soap to wash the cushion.
2. Use a wash cloth to wipe the cushion dry. Do not use detergents
or other solvents.
3. Do not machine wash or tumble dry.
4. Do not heat or steam autoclave.
COVER
1. Wash at 30˚C / 86˚F
2. Do not iron.
3. For single patient use only.
REPAIRING LEAKS
1. Make sure the cushion and all four corners are inflated completely.
2. Immerse the over-inflated cushion into water to locate the leak.
(You will see air bubbles rising from the area where the leak
is located.)
3. When the leak has been located, mark the leak.
4. Deflate the cushion.
5. Allow the cushion to dry completely.
6. Clean the area around the hole and let the surface dry.
7. Place a patch film over the hole.
8. Press the patch film down firmly to seal. Do not use the cushion
for at least 30 minutes.
9. When leak repair is complete, adjust the cushion to proper
inflation according to the user manual.
10. Repeat to test if there are any other leaks.
Patches do not change the performance of the cushion, but if your
patch is not successful in solving the problem, or there is more
serious damage, please contact your local provider.
Manufactured for: Dynarex Corporation
10 Glenshaw Street • Orangeburg, NY 10962
USA • www.dynarex.com
Made in China
Air Cushion
Reorder Nos. 10600, 10601, 10602, 10603, 10604
WARNING
WARNING
1. Before use, please read the user manual carefully.
2. Please pay attention to the positions contacting the cushion
bottom, which may produce sliding phenomenon.
3. Please don't use awls or other sharp tools to contact the cushion.
4. Fill the appropriate amount of air during use.
5. Keep away from fire and moist environments.
6. Please don't bend or fold the cushion in the gas filling state.
7. If any part of body touches the cushion base, too much air has
been released, which will reduce the therapeutic effect.
8. Please check the cushion every day to make sure the patient will
SYMBOL GLOSSARY
not bottom out.
For an explanation of symbols
used in Dynarex packaging, visit
dynarex.com/symbols.php
9. Please preserve the product and its accessories in
a dry environment.
R180109
Ever-Soft
10610, 10611, 10612, 10613, 10614
10670, 10671