For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The BP3L is intended for use in the electromagnetic environment specified below. The customer or the
user of the BP3L should assure that it is used in such an environment.
IMMUNITY test IEC 60601test level
Radiated RF
3 V/m 80 MHz to
IEC 61000-4-3
2.5 GHz
Table 3
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the BP3L, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
3 A/m
the transmitter.
Recommended separation distance:
80 MHz to 800 MHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
16
800 MHz to 2,5 GHz