Introduction 
The MR‐100™ Plus    Manual Resuscitator Is designed for use as an adjunct 
to artificial respiration and cardiopulmonary resuscitation. The Resuscitator 
can be used to ventilate the apnoeic patient and to augment ventilation 
and/or oxygen delivery to the spontaneously breathing patient.   
 
When connected to an oxygen source, the MR‐100 
Resuscitator can also deliver supplemental oxygen with the aid of an oxygen 
reservoir The oxygen concentration delivered by the MR‐100
Manual Resuscitator depends on the oxygen supply flow rate, tidal volume, 
ventilation frequency and operator technique.   
 
The Infant and Child models of the MR‐100™ Plus    Manual Resuscitator are 
fitted with pressure relief valves      which vent the delivered breath to the 
atmosphere if the pressure exceeds 40±5cmH20. The pressure relief valve 
can be overridden by the operator should a higher delivery pressure be 
required.   
 
The MR‐100™ Plus    Manual Resuscitator is intended for use by qualified 
medical and emergency personnel trained in pulmonary ventilation and 
advanced cardiac life support techniques. Users should read and understand 
the contents of this manual and demonstrate proficiency in the assembly, 
disassembly and use of this device prior to use. 
 
 
TM 
Plus    Manual 
TM
 Plus   
Warning And Cautions 
WARNING: Do not use the MR‐100"' Plus    Manual Resuscitator In toxic 
atmospheres. 
WARNING: Remove the oxygen reservoir and reservoir valve If 
supplemental oxygen Is not being administered. Failure to do so will 
affect the refill rate and maximum frequency capabilities. 
WARNING: Do not administer supplemental oxygen in the presence of 
open flames. 
WARNING: Do not use oil, grease or other hydrocarbon based 
substances on any part of the MR‐100
Supplemental oxygen, 
supplied under pressure, can combine with hydrocarbons and cause 
explosions. 
WARNING: This device is intended for use by qualified medical and 
emergency personnel trained in pulmonary ventilation and advanced 
cardiac life support techniques. 
WARNING: Always test the device in accordance with this manual after 
cleaning and disinfection or replacement of parts. 
WARNING: Do not use the pressure relief valve over rid mechanism for 
the child and infant resuscitator. 
CAUTION: Use only replacement parts intended for use on MR‐
100™Plus    Manual Resuscitators. 
CAUTION: Do not attempt to disassemble the pressure relief valve 
assembly. Disassembly will damage the component. 
 
 
 
 
TM 
Plus    Manual Resuscitator.