E. Apply approximately 1/8th
inch [3 mm] silicone
grease to the transducer
face. Place the trans-
ducer face on the pre-
pared area of pipe. Hold
the transducer parallel to
the pipe with the cable
pointing downstream of
the flow direction.
F. Press the ON/OFF key.
Wait at least 30 seconds
before recording a read-
ing. The UFX will auto-
matically turn itself off af-
ter 3 minutes.
G. Change units of measure
by pressing the FT/SEC -
M/SEC key.
STEP E
STEP F
5