GaleMed AR0097 Manual de instruções de utilização - Página 9

Procurar online ou descarregar pdf Manual de instruções de utilização para Equipamento médico GaleMed AR0097. GaleMed AR0097 12 páginas. Resuscitator silicone reservoir bag, 2l

GaleMed AR0097 Manual de instruções de utilização
4. Cleaning and Disinfection (Handling) 
 
This product is single patient multiple use, please dispose it after using it. 
 
4.1 Instruction for Cleaning before Use/Next Use 
This product is single patient multiple use, please dispose it after using it. 
 
4.2 Instruction for Disinfection/Sterilization before Use/Next Use 
 
This product is single patient multiple use, please dispose it after using it. 
 
4.3 Instruction for Testing before Use/Next Use 
 
Testing the silicone bag   
1. Remove the non‐rebreathing valve and the oxygen reservoir and valve (if fitted).   
2. Compress the silicone bag and occlude (block) the non‐rebreathing valve outlet.   
3. Release the bag. The bag should expand immediately and refill. If not, check that the intake valve at 
the base of the silicone bag is correctly assembled.   
4. While keeping the non‐rebreathing valve outlet blocked, compress the bag again. The bag should not 
compress easily. If this occurs, check that you are blocking the valve sufficiently, and that the intake valve 
at the base of the silicone bag is correctly assembled. 
 
Testing non‐rebreathing valve   
1. Connect the non‐rebreathing valve to the silicone bag. Connect the test lung to the outlet on the 
nonrebreathing valve.   
2. Compress and hold the bag. The non‐rebreathing (duckbill) valve inside the non‐rebreathing valve 
should open and the test lung should fill. If not, check the connection between the Resuscitator and the 
test lung, and check that the non‐rebreathing valve is correctly assembled.   
3. Release the bag. The non‐rebreathing (duckbill) valve should close and as the test lung deflates, gas 
should flow through the expiratory ports in the non‐rebreathing valve. If not, check that the non‐
rebreathing valve is correctly assembled.   
4. Ventilate the test lung for a minimum of 10 cycles to ensure that the Resuscitator is functioning 
correctly. Inspiration must occur when the silicone bag is compressed and exhalation when the bag is 
released. If not, check that the non‐rebreathing valve is correctly assembled. 
 
To check the function of the pressure relief valve (If available)   
Connect a 0‐100 cmH
bag. When the pressure relief valve activates, the manometer should read 35‐45 cmH2O. If not, check 
that the non‐rebreathing valve is correctly assembled and does not leak. If the pressure relief valve fails a 
further test, it must be replaced. Do not attempt to repair the pressure relief valve. 
 
 
O manometer to the patient outlet of the non‐rebreathing valve. Compress the 
2